MalfunctionJGS

ATELLICA CH A-LYTE INTEGRATED MULTISENSOR (IMT NA K CL)

Received Feb 12, 2025 · Event occurred Jan 7, 2025

Report 2432235-2025-00054 · MDR key 21368352

Device

Generic name

Electrode, Ion Specific, Sodium

Model number

N/A

Catalog number

11099315

Lot number

140018

Product problems

  • Low Test Results
  • Low Test Results

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

THE CUSTOMER REPORTED AN ERRONEOUSLY DEPRESSED SODIUM (NA) RESULT ON ONE (1) PATIENT SAMPLE USING AN A-LYTE INTEGRATED MULTISENSOR (IMT) OBTAINED ON AN ATELLICA CH 930 ANALYZER. THE ERRONEOUSLY DEPRESSED RESULT WAS NOT REPORTED TO THE PHYSICIAN. THE SAME SAMPLE WAS REPROCESSED ON AN ALTERNATE ATELLICA CH 930 ANALYZER. A HIGHER RESULT WAS OBTAINED, CONSIDERED CORRECT, AND REPORTED TO THE PHYSICIAN. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE ERRONEOUSLY DEPRESSED SODIUM (NA) RESULT.

Additional Manufacturer Narrative

AN OUTSIDE UNITED STATES (OUS) CUSTOMER CONTACTED THE SIEMENS HEALTHCARE DIAGNOSTICS REMOTE SERVICES CENTER (RSC) REGARDING AN ERRONEOUSLY DEPRESSED SODIUM (NA) RESULT ON ONE (1) PATIENT SAMPLE USING AN A-LYTE INTEGRATED MULTISENSOR (IMT) OBTAINED ON AN ATELLICA CH 930 ANALYZER. ANOTHER MDR IS BEING FILED FOR THE SAME EVENT. SIEMENS IS INVESTIGATING THE EVENT.

Additional Manufacturer Narrative

ADDITIONAL INFORMATION (12-MAR-2025): SIEMENS HEALTHCARE DIAGNOSTICS SERVICE OPERATIONS SUPPORT (SOS) CONCLUDED THE INVESTIGATION OF THE EVENT. SOS EVALUATED THE INFORMATION PROVIDED BY THE CUSTOMER AND THE AVAILABLE INSTRUMENT DATA. QUALITY CONTROL (QC) WAS WITHIN THE CUSTOMER¿S ACCEPTABLE RANGES. THE CAUSE OF THE EVENT IS UNKNOWN. THE CUSTOMER STATED THE ISSUE WAS RESOLVED AND NO OTHER DISCORDANT RESULTS WERE REPORTED. THE CUSTOMER IS OPERATIONAL. THE DEVICE IS PERFORMING WITHIN SPECIFICATIONS. A POTENTIAL PRODUCT ISSUE WAS NOT IDENTIFIED. NO FURTHER EVALUATION IS REQUIRED. SECTION H6 HAS BEEN UPDATED TO REFLECT THE SOS INVESTIGATION. INITIAL MDR 2432235-2025-00054 WAS FILED ON 12-FEB-2025. INITIAL MDR 2432235-2025-00055 WAS ALSO FILED ON 12-FEB-2025 FOR THE SAME EVENT.