Device

Generic name

Hemodialyzer, Re-use, High Flux

Catalog number

5M2733

Lot number

08B25B

Product problems

  • Fluid/Blood Leak

Patient

Not reported

  • No Information

Narrative

Description of Event or Problem

A CUSTOMER REPORTED TO BAXTER (B)(4) THAT FIVE UNITS OF CAHP-130 DIALYZER WERE NOTED LEAK OR BLOOD LEAK WITH ALARM ONLY AFTER TWO OR THREE TIMES OF REPROCESSING, WHICH MUCH LESS NORMAL REUSE TIMES (NORMAL REUSE 10 TIMES).

Additional Manufacturer Narrative

(B)(4). THE SAMPLE WAS RECEIVED AND EVALUATED. UPON VISUAL INSPECTION NO OBVIOUS DEFECTS WERE FOUND. THEN PERFORMED MANUAL LEAK TEST AND OBSERVED PRESSURE DECAY OF GREATER THAN 10 MMHG PER 60 SECONDS. FURTHER TESTING WAS DONE WITH MANUAL BUBBLE POINT TEST PER SAME SOP. PRESSURE DROPPED TO ZERO AND BUBBLES WERE SIGHTED AT THE VENOUS HEADER AND COMING OUT OF THE DIALYSATE PORT. THIS CONFIRMS THE COMPLAINT. THIS MDR IS BEING SUBMITTED AS PART OF BAXTER RENAL'S RETROSPECTIVE REVIEW & REMEDIATION PROJECT. THIS REPORT IS BEING FILED LATE TO FULFILL BAXTER'S COMMITMENT TO PERFORM A TWO YEAR RETROSPECTIVE REVIEW OF RENAL COMPLAINTS IN RESPONSE TO FDA-WARNING LETTER CHI-07-10.