inforMED
InjuryNBC

ACCESS B-TYPE NATRIURETIC PEPTIDE

Received Feb 12, 2025 · Event occurred Jan 14, 2025

Report 2122870-2025-00057 · MDR key 21365885

Device

Generic name

Test,natriuretic Peptide

Catalog number

98200

Lot number

439890

Product problems

  • High Test Results
  • High Test Results

Patient

NA · Unknown

  • Insufficient Information
  • Insufficient Information

Narrative

Additional Manufacturer Narrative

A1: THE FULL PATIENT IDENTIFIER IS (B)(4). A5: THE CUSTOMER DID NOT SUPPLY PATIENT DEMOGRAPHICS SUCH AS AGE, DATE OF BIRTH, SEX, WEIGHT, ETHNICITY, OR RACE. H3 AND H6: THE ACCESS BNP (ACCESS B-TYPE NATRIURETIC PEPTIDE) REAGENT WAS NOT RETURNED FOR EVALUATION. NO HARDWARE ERRORS OR ISSUES WITH OTHER ASSAYS WERE REPORTED IN CONJUNCTION WITH THIS EVENT. SYSTEM PERFORMANCE INDICATORS SUCH AS SYSTEM CHECK, CALIBRATION AND QUALITY CONTROL WERE PASSING WITHIN SPECIFICATIONS AT THE TIME OF THE EVENT. NO OTHER PATIENT RESULTS WERE CALLED INTO QUESTION. A FIELD APPLICATIONS SPECIALIST (FAS) REVIEWED THE SAMPLE HISTORY, CALIBRATION, QC, AND SYSTEM CHECK DATA FROM THE INSTRUMENT. THE FAS NOTED THAT THERE WAS NO EVIDENCE OF A HARDWARE MALFUNCTION AND NO FLAGS WERE GENERATED ON THE SAMPLE WHEN THE INSTRUMENT GENERATED THE RESULT. IN CONCLUSION, THE LIKELY CAUSE FOR THIS EVENT CANNOT BE DETERMINED WITH THE PROVIDED INFORMATION. THERE IS NO EVIDENCE TO REASONABLY SUGGEST THAT A MALFUNCTION OCCURRED IN CONJUNCTION WITH THIS EVENT.

Description of Event or Problem

ON (B)(6) 2025, THE CUSTOMER REPORTED ERRONEOUS ELEVATED ACCESS BNP (ACCESS B-TYPE NATRIURETIC PEPTIDE, PART NUMBER 98200, LOT NUMBER 439890) WERE GENERATED ON THE CUSTOMER¿S DXI 800 ACCESS IMMUNOASSAY W/SPOT B ANALYZER (PART NUMBER A71456, SERIAL NUMBER (B)(6) FOR ONE PATIENT. THE PATIENT¿S INITIAL SAMPLE GENERATED AN ELEVATED RESULT OF 774 PMOL/L ON (B)(6) 2025. THE PATIENT WAS REPORTED TO BE ADMITTED TO THE HOSPITAL WHERE ADDITIONAL TESTING WAS PERFORMED AND DID NOT SHOW ANY ABNORMALITY. THE ADDITIONAL TESTING PERFORMED IS NOT KNOWN AND WAS VERBALLY STATED TO BE ¿AN ECHO AND A CARDIOTHORACIC SCAN¿. THERE WAS NO FURTHER REPORT OF INJURY OF AFFECT TO THE PATIENT. AN ADDITIONAL SAMPLE WAS COLLECTED FROM THE PATIENT AND PRODUCED A RESULT OF 9 PMOL/L ON (B)(6) 2025. THE ORIGINAL SAMPLE WAS RE-RUN, ALTHOUGH PAST THE RECOMMENDED STABILITY) AND PRODUCED A RESULT OF 8 PMOL/L ON (B)(6) 2025. NO HARDWARE ERRORS OR ISSUES WITH OTHER ASSAYS WERE REPORTED IN CONJUNCTION WITH THIS EVENT. SYSTEM CHECK PASSED ON (B)(6) 2025. CALIBRATION PASSED ON (B)(6) 2024 WITH REAGENT LOT 439890 AND CALIBRATOR LOT 439218. QUALITY CONTROL (QC) WAS PASSING WITHIN THE LABORATORY¿S ESTABLISHED RANGES. NO OTHER PATIENT RESULTS WERE CALLED INTO QUESTION. NO ISSUES WITH SAMPLE INTEGRITY WERE REPORTED BY THE CUSTOMER.