Additional Manufacturer Narrative

THE THIRD-PARTY REPAIR SERVICEMAN INSPECTED THE INSTRUMENT AND FOUND THE BRAKE RESISTOR TO BE THE SOURCE OF THE ISSUE. THE SERVICEMAN ORDERED THE REPLACEMENT BRAKE RESISTOR ASSEMBLY AS WELL AS A POWER CONTROL BOARD ASSEMBLY AND TACHOMETER ASSEMBLY TO RESOLVE THE ISSUE. BEC INTERNAL IDENTIFIER -(B)(4).

Description of Event or Problem

THE CUSTOMER REPORTED THAT THEIR AVANTI J-15R CENTRIFUGE WAS OVERHEATING WHILE SITTING IDLE AND FELT HOT TO THE TOUCH. THE REPORTED ISSUE INITTIALLY STATED WITH A D505 ERROR MESSAGE. THERE WERE NO REPORTS OF INJURIES, DEATHS, OR DELAYS/CHANGES IN TREATMENT OR DIAGNOSIS RELATED TO THE REPORTED ISSUE. THERE WERE NO REPORTS OF HARM TO A PATIENT OR OPERATOR.