MalfunctionFOH
TEMP PUMP TEMPERATURE
Received Jun 14, 2011 · Event occurred May 19, 2011
Report 1313850-2011-00140 · MDR key 2135238
Device
Product problems
- Break
- Failure to Pump
- Material Protrusion/Extrusion
Patient
Not reported
- No Patient Involvement
Narrative
Description of Event or Problem
IT WAS REPORTED THAT THE UNIT WOULD NOT PUMP WATER. EVAL FOUND THAT THE POWER CORD HAD BEEN DAMAGED AND GROUND WIRE WAS PARTIALLY EXPOSED. NO PT INVOLVEMENT OR ADVERSE CONSEQUENCES WERE REPORTED.