Narrative
Description of Event or Problem
IT WAS REPORTED THAT THE PAD APPEARED TO HAVE A BLOCKAGE AND IT WOULD NOT ALLOW WATER TO FLOW THROUGH THE ENTIRE PAD. NO PT INVOLVEMENT OR ADVERSE CONSEQUENCES WERE REPORTED. CUSTOMER STATED THAT THE PAD APPEARED TO HAVE A BLOCKAGE IN. IT WOULDN'T LET WATER FLOW THROUGH THE WHOLE PAD.
Additional Manufacturer Narrative
RESULT: INTERNAL BUTTON WELDS HAVE PEELED, CAUSING THE MATERIAL TO STRETCH AND FORM HOLES. CONCLUSION: CUSTOMER REQUESTED CREDIT FOR PAD.