inforMED
MalfunctionLJX

CHEMSTRIP 10 MD URINE TEST STRIPS

Received Feb 10, 2025 · Event occurred Jan 16, 2025

Report 1823260-2025-00386 · MDR key 21347047

Add to watch list
Product code (LJX)
Manufacturer (Roche Diagnostics International Ltd.)
Model (D051188)

Device

Generic name

Urine Test Strips

Manufacturer

Roche Diagnostics

Model number

D051188

Catalog number

03260763160

Lot number

80097404

Product problems

  • Incorrect Measurement
  • Device Difficult to Setup or Prepare
  • Incorrect Measurement
  • Device Difficult to Setup or Prepare

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

SECTIONS D1-D4, G1, AND G4 WERE UPDATED. THE RETENTION MATERIAL FOR LOT 80097400 WAS MEASURED ON AN IU URISYS 1100 ANALYZER EQUIPPED WITH SW-V. 6.71 AND BY VISUAL READING WITH 0-NATIVE-URINE AND A PROTEIN-DILUTION-SERIES ACCORDING TO TESTING-PLAN. THE TESTING SHOWED NO ABNORMAL RESULTS. THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM.

Additional Manufacturer Narrative

THE URINE TEST STRIP LOT NUMBER IS 1080097404 WITH AN EXPIRATION DATE OF 30-NOV-2025. IT WAS REPORTED THE CUSTOMER RECEIVED AN E-5 ALARM DURING A CALIBRATION ISSUE. THE TEST STRIPS ARE NOT AVAILABLE FOR INVESTIGATION AS THEY HAVE BEEN DISCARDED. THE INVESTIGATION IS ONGOING.

Description of Event or Problem

THERE WAS AN ALLEGATION OF QUESTIONABLE NEGATIVE PROTEIN RESULTS FOR APPROXIMATELY 6 PATIENT SAMPLES ON A URISYS 1100. IT WAS ALLEGED THAT THE VISUAL RESULTS WERE 2 COLOR BLOCKS OUT (POSITIVE).