inforMED
MalfunctionITQ

DONJOY

Received Feb 7, 2025 · Event occurred Jan 13, 2025

Report 9616086-2025-00006 · MDR key 21339784

Device

Generic name

X-rom Post-op Knee Brace

Manufacturer

Djo Llc

Catalog number

11-2181-9

Lot number

UNLKNOWN

Product problems

  • Mechanical Problem
  • Unintended System Motion
  • No Apparent Adverse Event
  • Mechanical Problem
  • Unintended System Motion
  • No Apparent Adverse Event

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

IT WAS REPORTED THAT THE LOCK BUTTON IS BROKEN, COULD LOCK IT BUT ONCE SHE STOOD UP IT WOULD AUTOMATICALLY UNLOCK. THE BRACE HAS NOT BEEN RETURNED FOR EVALUATION. THE ROOT CAUSE OF THE COMPLAINT IS DAMAGED COMPONENT BASED ON OTHER DEVICES THAT HAVE BEEN EVALUATED. THE DEVICES LOCK BUTTON FAILED DURING MANUFACTURING TESTING CONFIRMED TO BE THE SAME MALFUNCTION FOUND IN REPORT NUMBER 9616086-2023-00007. ADDITIONAL REPORTING ON THIS EVENT WILL BE PROVIDED AS A SUPPLEMENTAL REPORT TO THIS DOCUMENT IF IT BECOMES AVAILABLE.

Description of Event or Problem

IT WAS REPORTED THAT THE LOCK BUTTON IS BROKEN, COULD LOCK IT BUT ONCE SHE STOOD UP IT WOULD AUTOMATICALLY UNLOCK.