inforMED
MalfunctionFLE

AMSCO

Received Feb 7, 2025 · Event occurred Jan 9, 2025

Report 3005899764-2025-00011 · MDR key 21336211

Device

Generic name

Amsco 600 Steam Sterilizer

Catalog number

EF1201242

Product problems

  • Leak/Splash
  • Leak/Splash

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

A STERIS SERVICE TECHNICIAN ARRIVED ON SITE TO INSPECT THE AMSCO 600 STERILIZER AND FOUND IT NOT OPERATIONAL, UPON FURTHER INSPECTION THE TECHNICIAN FOUND THAT THE LEAK WAS COMING FROM THE SIGHT GLASS DUE TO A DAMAGED GASKET. TO RESOLVE THE ISSUE, THE TECHNICIAN REPAIRED THE SIGHT GLASS WITH A REPAIR KIT, TESTED THE UNIT, FOUND IT TO BE OPERATING TO SPECIFICATION, AND RETURNED IT TO SERVICE. NO ADDITIONAL ISSUES HAVE BEEN REPORTED.

Description of Event or Problem

THE USER FACILITY REPORTED THAT THEIR AMSCO 600 STERILIZER WAS LEAKING WATER ONTO THE FLOOR. NO REPORT OF INJURY.