EXOGEN ULTRASOUND HEALING SYSTEM

Received Mar 8, 1999 · Event occurred Feb 3, 1999

Report MW1015854 · MDR key 213347

Device

Generic name

External Ultrasound Healing System

Manufacturer

Exogen, Inc.

Model number

THM MODEL 2000

Catalog number

NA

Lot number

NA

Product problems

  • Component(s), worn
  • Design/structure problem
  • Electrical /Electronic Property Problem
  • Repair
  • Replace
  • Inadequate Service
  • Other (for use when an appropriate device code cannot be identified)

Patient

48 YR

  • Therapy/non-surgical treatment/delayed/prolonged

Narrative

Description of Event or Problem

DR STATES SYSTEM WORKS VERY WELL FOR PTS WHEN IT WORKS, BUT AFTER 50-60 USES THE HEAD WEARS OUT OR GETS GEL TRAPPED INSIDE, LEADING TO SHORT CIRCUITING. IN HIS CASE, HE HAS RETURNED THE DEVICE FOR REPAIR OR REPLACEMENT. THE CO CHECKS IT OUT AND RETURNS IT TO HIM SAYING THAT IT IS FUNCTIONING. HE FINDS THAT IT ISN'T AND LOSES MORE TIME BETWEEN TREATMENTS. RPTR FEELS THAT THE CO SHOULD PROVIDE BETTER SERVICE TO THEIR PTS, BY SENDING NEW UNITS IN EXCHANGE FOR FAULTY ONES AND PERHAPS IMPLEMENT AN AUTOMATIC REPLACEMENT IF THE HEAD AFTER 50-60 TREATMENTS. THESE STEPS, THE RPTR FEELS, WOULD PREVENT PTS FROM LOSING TREATMENT EFFECTS WAITING FOR REPAIR OR REPLACEMENT. IF THESE STEPS ARE NOT TAKEN, THE RPTR FEELS THAT THE DEVICE SHOULD BE RECALLED FOR REDESIGN.