inforMED
MalfunctionOCZ

FORCEPS, 5 FR., 60 CM

Received Feb 5, 2025 · Event occurred Apr 10, 2024

Report 9610617-2025-00139 · MDR key 21318175

Device

Generic name

Forceps, 5 Fr., 60 Cm

Model number

27425F

Catalog number

27425F

Lot number

WN04

Product problems

  • Detachment of Device or Device Component
  • Detachment of Device or Device Component

Patient

NA · Unknown

  • Insufficient Information
  • No Clinical Signs, Symptoms or Conditions
  • Insufficient Information
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

ADDITIONAL INFORMATION IS PROVIDED IN SECTION B5 AND SECTION H10. THE EVENT IS FILED UNDER INTERNAL KARL STORZ COMPLAINT ID HAS BEEN CHANGED TO (B)(4).

Description of Event or Problem

NEW INFORMATION WAS PROVIDED THAT STATES TWO PATIENTS WERE INVOLVED IN THIS EVENT.

Additional Manufacturer Narrative

THE DEVICE IN QUESTION WAS RETURNED TO THE MANUFACTURER. THIS INFORMATION IS REFLECTED IN SECTION D9 IT WAS RETURNED ON DECEMBER 23,2025. THE EVALUATION IS ANTICIPATED, BUT NOT YET BEGUN. THE INVESTIGATION WILL BE PERFORMED BY A DESIGNATED KARL STORZ EMPLOYEE. THE EVENT IS FILED UNDER INTERNAL KARL STORZ COMPLAINT ID: (B)(4).

Description of Event or Problem

IT WAS REPORTED DURING THE OPERATION, WHEN GRABBING THE STONE, THE CONNECTION OF THE BRANCHES WITH THRUST WAS DAMAGED. THE REPLACEMENT WAS USED IN 5-10 MINUTES. THE PROCEDURE WAS COMPLETED WITHOUT ANY PATIENT INJURY OR HARM, THERE WAS NO PATIENT INJURY. NO PATIENT INVOLVEMENT AND PROCEDURE DELAY LESS THAN 15 MINS. CROSS REFERENCE MDR: 9610617-2025-00137.

Additional Manufacturer Narrative

THE PLIERS SHOW SEVERE CORROSION AT THE PROXIMAL CONNECTION POINT AND IN THE JAW AREA, WHICH MAY INDICATE IMPROPER REPROCESSING. ADDITIONALLY, EXCESSIVE MECHANICAL STRAIN ON THE JAWS COULD HAVE LED TO BREAKAGE OR LOOSENING AT THE CONNECTION POINT. THE INVESTIGATION DID NOT REVEAL ANY ROOT CAUSE RELATED TO THE LABELLING, THE DEVICE, OR ITS HISTORY. THE EVENT IS FILED UNDER INTERNAL KARL STORZ COMPLAINT ID: (B)(4).