ELECSYS FT3 III
Received Feb 4, 2025 · Event occurred Jan 13, 2025
Report 1823260-2025-00332 · MDR key 21306974
Device
Generic name
Total Triiodothyronine
Manufacturer
Roche DiagnosticsModel number
FT3 IIICatalog number
09005811190
Lot number
781598
Product problems
- High Test Results
- High Test Results
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
MEDWATCH FIELDS D1, D2A, D4 MODEL NUMBER, LOT NUMBER, EXPIRATION DATE AND CATALOG NUMBER, UDI NUMBER, AND G4 PMA / 510K WERE UPDATED.
Additional Manufacturer Narrative
THE E801 ANALYZER SERIAL NUMBER WAS (B)(6). THE COMPETITOR METHOD IS SNIBE. THE SAMPLE WAS REQUESTED FOR INVESTIGATION.
Description of Event or Problem
THE INITIAL REPORTER QUESTIONED THYROID RESULTS FOR 1 PATIENT ON A COBAS E 801 ANALYTICAL UNIT COMPARED TO A COMPETITOR METHOD. OF THE DATA PROVIDED, THE RESULTS FOR FT3, T4, T3, ANTI-TG, AND ANTI-TPO WERE DISCREPANT. REFER TO THE ATTACHED DATA FOR THE PATIENT RESULTS. THIS MEDWATCH WILL COVER FT3. REFER TO MEDWATCH WITH A1 PATIENT IDENTIFIER (B)(6) FOR INFORMATION ON THE T4 RESULTS, MEDWATCH WITH A1 PATIENT IDENTIFIER (B)(6) FOR INFORMATION ON THE T3 RESULTS, MEDWATCH WITH A1 PATIENT IDENTIFIER (B)(6) FOR INFORMATION ON THE ANTI-TG RESULTS AND MEDWATCH WITH A1 PATIENT IDENTIFIER (B)(6) FOR INFORMATION ON THE ANTI-TPO RESULTS.
Additional Manufacturer Narrative
THE SAMPLE WAS SUBMITTED FOR INVESTIGATION. UPON FURTHER INVESTIGATION OF THE PATIENT SAMPLE, AN INTERFERENT AGAINST THE STREPTAVIDIN COMPONENT OF THE REAGENT WAS CONFIRMED. THIS INTERFERENCE IS COVERED IN PRODUCT LABELING: "IN RARE CASES, INTERFERENCE DUE TO EXTREMELY HIGH TITERS OF ANTIBODIES TO ANALYTE-SPECIFIC ANTIBODIES, STREPTAVIDIN OR RUTHENIUM CAN OCCUR. THESE EFFECTS ARE MINIMIZED BY SUITABLE TEST DESIGN. FOR DIAGNOSTIC PURPOSES, THE RESULTS SHOULD ALWAYS BE ASSESSED IN CONJUNCTION WITH THE PATIENT¿S MEDICAL HISTORY, CLINICAL EXAMINATION AND OTHER FINDINGS." THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM.