inforMED
InjuryNRY

072 ASPIRATION SYSTEM

Received Jan 31, 2025 · Event occurred Dec 31, 2023

Report 3015614180-2025-00002 · MDR key 21284855

Device

Generic name

Catheter, Thrombus Retriever,

Model number

FG 00806-02

Catalog number

AP6072-132

Lot number

FG231222A-01

Product problems

  • Use of Device Problem
  • Use of Device Problem

Patient

51 YR · Female

  • Extravasation
  • Hemorrhage/Blood Loss/Bleeding
  • Extravasation
  • Hemorrhage/Blood Loss/Bleeding

Narrative

Description of Event or Problem

RETROSPECTIVE FILING FOR AN EVENT THAT OCURED 5/2/2024. THE CASE WAS REPORTED BY Q'APEL MEDICAL'S CHIEF MARKETING OFFICER WHO WAS PRESENT FOR THE CASE AND REPORTED THE FINDINGS. THE INITIAL REVIEW BY Q'APEL MEDICAL'S MEDICAL PHYSICAN AND "QUALIFIED PERSON" DETERMINED THAT THE Q'APEL DEVICE DID NOT CONTRIBUTE TO THE INJURY AND THEREFORE, THE EVENT WAS NOT REPORTABLE. NEW INFORMATION WAS OBTAINED BY Q'APEL MEDICAL FROM THE OPERATOR OF THE DEVICE ON 1/15/2025 WHICH STATED: "THERE IS NO FLUORO LOOP OF THE 2ND PASS HIPPO NAVIGATING UP TO THE CLOT. THERE IS ONE OF THE FIRST PASS AND IT NAVIGATED SMOOTHLY WITH NO WIRE ISSUES. I CAN NOT IMAGINE HOW THAT AMOUNT OF EXTRAVASATION OCCURRED FROM A MICROWIRE PERFORATION. " THE NEW INFORMATION CONCLUDED THAT THE Q'APEL MEDICAL DEVICE LIKELY CONTRINUTED TO THE INJURY THAT REQUIRED MEDICAL INTERVENTION. CASE DATE:(B)(6) 2024. 78 YO [YEAR OLD] FEMALE PATIENT. STROKE CASE R M1 OCCLUSION. 3.5 SINCE LKW. NO TNK OR TPA ADMINISTERED DUE TO A FALL. WALRUS BGC USED AS GUIDE. 1 PASS WITH HIPPO CHEEATH WITH AN ARISTOTLE 18 GW. RESULT WAS TICI 0. 2ND PASS WITH HIPPO CHEETAH AND ARISTOTLE 18 GW. ON RUN POST ASPIRATION, VESSEL RUPTURE WITH CONTRAST EXTRAVASATION VITALS REMAINED STABLE FOR PATIENT AND NO LOSS OF PRESSURE. SCEPTER BALLOON PLACED AT SITE TO ARREST BLEED FOR 5 MIN. BLEEDING CONTROLLED. NO OTHER INTERVENTION TO CONTROL THIS REQUIRED. 3RD PASS - ZOOM 55. TICI 0. 4TH PASS - ZOOM 55 TICI 0. 5TH PASS - VELOCITY MC AND TIGERTRIEVER 32MM DEPLOYED FOR 2 MINS - TICI 0. 6TH PASS - ZOOM 55 - TICI 0.