MalfunctionLMB

IMAGE MANAGEMENT SYSTEM 660HD-E

Received Jan 30, 2025 · Event occurred Jan 14, 2025

Report 1643264-2025-00033 · MDR key 21277379

Device

Generic name

Device, Digital Image Storage, Radiological

Catalog number

72204668

Product problems

  • Computer Software Problem
  • Computer Software Problem

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

H3, H6: THE REPORTED DEVICE WAS RECEIVED FOR EVALUATION. A VISUAL INSPECTION REVEALED NO DEFECTS. A FUNCTIONAL EVALUATION WAS PERFORMED ON THE RETURNED DEVICE AND FOUND IT STARTS WITH THE MESSAGE: REBOOT AND SELECT PROPER BOOT. A REVIEW OF DEVICE RECORDS SHOWED THERE WERE NO INDICATIONS TO SUGGEST THAT THE PRODUCT DID NOT MEET MANUFACTURING SPECIFICATIONS UPON RELEASE FOR DISTRIBUTION. A COMPLAINT HISTORY REVIEW FOUND SIMILAR REPORTED EVENTS. A RISK MANAGEMENT REVIEW FOUND THAT THE REPORTED FAILURE AND/OR HARM WAS DOCUMENTED APPROPRIATELY, AND THERE WERE NO INDICATIONS TO SUGGEST THE ANTICIPATED RISK IS NOT ADEQUATE. THE ROOT CAUSE HAS BEEN ASSOCIATED WITH A COMPONENT FAILURE. FACTORS THAT COULD HAVE CONTRIBUTED TO THE FAILURE INCLUDE IMPROPER SHUTDOWNS, OR RECENT HARDWARE OR SOFTWARE CHANGES. BASED ON THIS INVESTIGATION, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED.

Additional Manufacturer Narrative

H10: INTERNAL COMPLAINT REFERENCE: (B)(4). H3, H6: THIS COMPLAINT WAS OPENED BY SMITH+NEPHEW TO DOCUMENT A PRODUCT PROBLEM ASSOCIATED WITH A SMITH+NEPHEW DEVICE. THE REPORTED PROBLEM RELATES TO KNOWN INHERENT DEVICE AND/OR PROCEDURAL RISKS THAT ARE APPROPRIATELY DOCUMENTED IN OUR RISK FILES. SMITH+NEPHEW WILL CONTINUE TO MONITOR TRENDS IN ACCORDANCE WITH OUR POST-MARKET SURVEILLANCE PROCESS AND TAKE NECESSARY ACTION AS REQUIRED IF ANTICIPATED SEVERITY AND/OR OCCURRENCE RATES ARE EXCEEDED. SMITH+NEPHEW HAS NO REASON TO SUSPECT THAT THE PRODUCT FAILED TO MEET ANY SPECIFICATIONS AT THE TIME OF MANUFACTURE. BASED ON OUR REVIEW OF ALL CURRENTLY AVAILABLE INFORMATION, WE ARE UNABLE TO IDENTIFY A DEFINITIVE ROOT CAUSE. HOWEVER, AS THE USE OF OUR PRODUCT CANNOT BE EXCLUDED AS A POTENTIAL CAUSE OR CONTRIBUTORY FACTOR TO THE REPORTED ISSUE, WE ARE CONSERVATIVELY SUBMITTING THIS REPORT IN ACCORDANCE WITH APPLICABLE REGULATIONS. IF ADDITIONAL INVESTIGATIVE FINDINGS OR INFORMATION BECOMES AVAILABLE THAT ALTERS THE CONCLUSIONS OF THIS REPORT, A FOLLOW-UP REPORT WILL BE SUBMITTED AS REQUIRED.

Description of Event or Problem

IT WAS REPORTED THAT DURING AN ARTHROSCOPY, THE IMAGE MANAGEMENT SYSTEM WAS NOT STARTING CORRECTLY AND REQUIRED BOOT SOFTWARE TO START. IT WAS UNUSABLE. THE PROCEDURE WAS COMPLETED WITH A SURGICAL DELAY GREATER THAN 30 MINUTES USING A SMITH AND NEPHEW BACK UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.