NICVIEW 2.0 CAMERA
Received Jan 29, 2025 · Event occurred Jan 23, 2025
Report 9612330-2025-00004 · MDR key 21264167
Device
Product problems
- Excessive Heating
- Excessive Heating
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
INITIAL REPORT REF NATUS COMPLAINT # (B)(4). NO RELATED CAPAS. A REPLACEMENT NICE VIEW 2.0 WAS SENT TO THE CUSTOMER, AND THEY WERE REQUESTED TO RETURN THE DEFECTIVE DEVICE.
Description of Event or Problem
CUSTOMER REPORTS THAT CAMERA WAS USED FOR A PATIENT WHO WAS ON AN OPEN WARMER. THE CAMERA WAS NOT NEAR THE HEATING ELEMENT, NOR WAS IT DIRECTLY UNDER THE HEATING ELEMENT BUT SOMEHOW, IT "MELTED". NO ADVERSE EFFECTS TO THE USERS OR PATIENTS.
Description of Event or Problem
CUSTOMER REPORTS THAT CAMERA WAS USED FOR A PATIENT WHO WAS ON AN OPEN WARMER. THE CAMERA WAS NOT NEAR THE HEATING ELEMENT, NOR WAS IT DIRECTLY UNDER THE HEATING ELEMENT BUT SOMEHOW, IT "MELTED". NO ADVERSE EFFECTS TO THE USERS OR PATIENTS.
Additional Manufacturer Narrative
FOLLOW UP REPORT REF NATUS COMPLAINT #(B)(4). INITIAL REPORT INDICATED THE CAMERA WAS A SUFFICIENT DISCTANCE FROM A HEATING LAMP BUT ONSITE TECH CONFIRMED THAT "CAMERA GOT TOO CLOSE TO LAMP AND MELTED". THIS IS USER ERROR AND NOT A PRODUCT MALFUCTION. RISK MANAGEMENT FILE REVIEW: (B)(4): NICVIEW 2 RISK ANALYSIS RISK MANAGEMENT FILE REVISION: J CAUE: LOSS OR INTERMITTENT VIDEO STREAMING FUNCTION HARM: SYSTEM NOT AVAILABLE FOR USE CREATING INCONVENIENCE OR ANNOYANCE TO OPERATOR OR USERS. RISK HAZARD ID LINE ITEM: 14.2 RISK SEVERITY: 0 RESIDUAL RISK LEVEL: LOW. RISK LEVEL DEEMED LOW AND ACCEPTABLE.
Remedial action
- Replace