MINI JOINT DISTRACTOR/COMPRESSOR
Received Jan 29, 2025 · Event occurred Jan 6, 2025
Report 1220246-2025-00098 · MDR key 21264012
Device
Generic name
Orthopedic Manual Surg Instr
Manufacturer
Arthrex, Inc.Model number
MINI JOINT DISTRACTOR/COMPRESSORCatalog number
AR-8970JD
Lot number
6402180814
Product problems
- Physical Resistance/Sticking
- Physical Resistance/Sticking
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
INVESTIGATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE PROVIDED UPON AVAILABILITY OF ADDITIONAL INFORMATION.
Description of Event or Problem
ON 01/06/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-8970JD MINI JOINT DISTRACTOR/COMPRESSORS SWITCH TO FLIP THE DISTRACTOR FROM COMPRESSION-NEUTRAL-DISTRACTION POSITIONS, IS DIFFICULT TO ENGAGE. THIS OCCURRED DURING A CASE AND CAUSE NO HARM ON THE PATIENT.
Additional Manufacturer Narrative
COMPLAINT ALLEGATION IS CONFIRMED. VISUAL EVALUATION NOTED THAT THE KNOB HAS AN ABRASION MARK ON THE TOP OF THE THREE EDGES. THE PART HAS DISCOLORATION ON THE ARM RACK OF THE MINI JOINT DISTRACTOR/COMPRESSOR. FUNCTIONAL TESTING WAS PERFORMED THE KNOB WAS FOUND THAT FLIPPED EASILY AND DID NOT SEAT ON THE NEUTRAL POSITION. THE MOST LIKELY REASON FOR THE REPORTED FAILURE IS THE WEAR AND TEAR DAMAGE INCURRED OVER REPEATED USAGE IN THE FIELD. MANUFACTURING DATE 2020.