inforMED
MalfunctionFAM

PENTAX

Received Jan 28, 2025 · Event occurred Jan 9, 2025

Report 9610877-2025-50517 · MDR key 21249809

Device

Generic name

Video Sigmoidoscope 4.2c 12.8t 700l Fwj

Model number

ES-3870K

Product problems

  • Fluid/Blood Leak
  • Fluid/Blood Leak

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(FLUID DAMAGE).

Additional Manufacturer Narrative

THE PRODUCT WAS RETURNED TO PENTAX MEDICAL FOR REPAIR. OUR TECHNICIAN CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE CCD MODULE WITH DRIVE PCB FLUID DAMAGE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE FLUID DAMAGE FROM THE CCD MODULE WITH DRIVE PCB. IN ADDITION, OUR TECHNICIAN CONFIRMED THAT THE INSERTION FLEXIBLE TUBE LEAK, AND THE INSERTION FLEXIBLE TUBE CUT; HOWEVER, THESE DEFECTS ARE NOT THE MAIN CAUSE, AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT. BASED ON THE TECHNICAL REPORT ""HR-RPT-0586(IMAGE FAILURE)"" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.