DIMENSION PHOSPHORUS FLEX REAGENT CARTRIDGE
Received Jan 27, 2025 · Event occurred Sep 28, 2024
Report 2517506-2025-00014 · MDR key 21242683
Device
Generic name
Inorganic Phosphorus
Manufacturer
Siemens Healthcare Diagnostics Inc.Model number
N/ACatalog number
10720277
Lot number
BA5119
Product problems
- High Test Results
- High Test Results
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
ADDITIONAL INFORMATION (10-FEB-2025): SIEMENS SERVICE OPERATIONS SUPPORT (SOS) CONCLUDED THE INVESTIGATION OF THE EVENT. SOS REVIEWED THE INFORMATION PROVIDED BY THE CUSTOMER. NO REAGENT ISSUES WERE IDENTIFIED. A SIEMENS CUSTOMER SERVICE ENGINEER (CSE) WAS DISPATCHED TO THE CUSTOMER'S SITE AND INSPECTED THE INSTRUMENT. A WHITE PIECE OF MATERIAL WAS FOUND ON THE SYRINGE TIP ON THE INSTRUMENT. NO DISCORDANT RESULTS WERE REPORTED AFTER THE SYRINGE TIP WAS REPLACED. A POTENTIAL PRODUCT ISSUE WAS NOT IDENTIFIED. THE CUSTOMER IS OPERATIONAL. THE DEVICE IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION IS REQUIRED. IN SECTION H6, THE INVESTIGATION FINDING AND INVESTIGATION CONCLUSION CODES WERE UPDATED. INITIAL MDR 2517506-2025-00014 WAS FILED ON 27-JAN-2025.
Additional Manufacturer Narrative
A UNITED STATES (US) CUSTOMER CONTACTED THE SIEMENS HEALTHCARE DIAGNOSTICS REMOTE SERVICES CENTER (RSC) REGARDING AN ERRONEOUSLY ELEVATED PHOSPHORUS (PHOS) RESULT ON A DIMENSION EXL 200 INTEGRATED CHEMISTRY SYSTEM. SIEMENS HEALTHCARE DIAGNOSTICS IS INVESTIGATING THE EVENT.
Description of Event or Problem
THE CUSTOMER REPORTED TO SIEMENS THEY OBTAINED AN ERRONEOUSLY ELEVATED PHOSPHORUS (PHOS) RESULT ON A DIMENSION EXL 200 INTEGRATED CHEMISTRY SYSTEM. THE ERRONEOUSLY ELEVATED RESULT WAS REPORTED TO THE PHYSICIAN AND NOT QUESTIONED. THE SAME SAMPLE WAS REPROCESSED ON THE ORIGINAL DIMENSION EXL 200 INTEGRATED CHEMISTRY SYSTEM. A LOWER RESULT WAS OBTAINED, CONSIDERED CORRECT, AND REPORTED TO THE PHYSICIAN. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE ERRONEOUSLY ELEVATED PHOSPHORUS (PHOS) RESULT.