GRAFTMASTER
Received Jan 22, 2025 · Event occurred Dec 3, 2024
Report 2024168-2025-00786 · MDR key 21211778
Device
Generic name
Coronary Stent Delivery System
Manufacturer
Abbott Vascular Inc.Model number
1012817-26CCatalog number
1012817-26C
Lot number
4051541
Product problems
- Failure to Advance
- Failure to Advance
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A PERFORATION ON A DE NOVO, HEAVILY CALCIFIED, HEAVILY TORTUOUS LEFT CIRCUMFLEX ARTERY. THE 2.80X26MM GRAFTMASTER COVERED STENT FAILED TO CROSS AFTER SEVERAL ATTEMPTS DUE TO ANATOMY. A NEW GRAFTMASTER COVERED STENT WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT SEQUELA AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Additional Manufacturer Narrative
MANUFACTURER'S INVESTIGATION IS STILL PENDING AT THIS TIME. RESULTS AND CONCLUSIONS WILL BE PROVIDED IN THE FINAL REPORT.
Additional Manufacturer Narrative
A VISUAL AND DIMENSIONAL INSPECTION WAS PERFORMED ON THE RETURNED DEVICE. THE REPORTED FAILURE TO ADVANCE COULD NOT BE EVALUATED AS THE EXACT ANATOMICAL CONDITIONS ENCOUNTERED BY THE DEVICE USED DURING THE PROCEDURE COULD NOT BE REPLICATED IN THE TEST LABORATORY. PRODUCTION RECORD AND CORRECTIVE AND PREVENTATIVE ACTIONS (CAPA) REVIEWS WERE PERFORMED AND REVEALED NO INDICATION OF A PRODUCT QUALITY ISSUE. ADDITIONALLY, A QUERY OF THE COMPLAINT HANDLING DATABASE FOR THE REPORTED LOT REVEALED THERE IS NO INDICATION OF A LOT SPECIFIC ISSUE. BASED ON THE INFORMATION RECEIVED AND ANALYSIS OF THE RETURNED DEVICE, THE INVESTIGATION DETERMINED THAT THE REPORTED FAILURE TO ADVANCE APPEARS TO BE RELATED TO CIRCUMSTANCES OF THE PROCEDURE, AS IT IS LIKELY THE DEVICE INTERACTED WITH THE HEAVILY CALCIFIED, HEAVILY TORTUOUS LESION DURING ADVANCEMENT, RESULTING IN THE REPORTED FAILURE TO ADVANCE. BASED ON THE RESULTS OF THE COMPLAINT INVESTIGATION, THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO THE DESIGN, MANUFACTURE, OR LABELING OF THE DEVICE.