BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE
Received Jan 21, 2025 · Event occurred Dec 25, 2024
Report 1917413-2024-01403 · MDR key 21200374
Device
Generic name
Lymphocyte Separation Medium
Manufacturer
Becton, Dickinson & Co. (broken Bow)Catalog number
362760
Lot number
4197918
Product problems
- Device Contamination with Chemical or Other Material
- Device Contamination with Chemical or Other Material
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
E1: INITIAL REPORTER FACILITY NAME: (B)(6). INVESTIGATION SUMMARY: BD RECEIVED SIX PHOTOS FOR INVESTIGATION. THE EVALUATION OF THESE PHOTOS REVEALED FLOATING REDDISH-BROWN STRANDS IN THE LIQUID DENSITY MEDIA (LDM) LAYER OF THE TUBE. THESE STRANDS ARE COMPOSED OF RESIDUAL SILICONE COATING AND TRACE METALS FROM THE LDM, WHICH ARE A RESULT OF THE MANUFACTURING PROCESS AND ARE NOT CONSIDERED FOREIGN MATTER AS THEY ARE INTRINSIC TO THE COMPONENTS OF THE TUBE. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THIS COMPLAINT HAS NOT BEEN CONFIRMED FOR THE INDICATED FAILURE MODE: FOREIGN MATTER - UNKNOWN. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR THE IDENTIFICATION OF EMERGING TRENDS.
Description of Event or Problem
IT WAS REPORTED PRIOR TO USE OF BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE 110 TUBES CONTAINED FOREIGN MATTER(FM). THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.