IMRIS ORT300 OPERATING ROOM TABLE

Received Jan 17, 2025 · Event occurred Dec 20, 2024

Report 3010326005-2025-00002 · MDR key 21185499

Device

Generic name

Operating Room Table

Model number

ORT300

Catalog number

114093-000

Product problems

  • Unintended Movement
  • Unintended Movement

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

THE CUSTOMER REPORTED THAT DURING OPERATING ROOM PREPARATION, THEY WERE NOT ABLE TO LOCK THE ROTATIONAL POSITION OF THE ORT300 REMOVABLE OPERATING ROOM TABLE. THERE WAS NO PATIENT INVOLVEMENT. AN ALTERNATE ORT300 REMOVABLE OPERATING ROOM TABLE WAS UTILIZED IN THE SUBSEQUENT PROCEDURE.

Additional Manufacturer Narrative

AN IMRIS SERIVCE ENGINEER ASSESSED THE OPERATING ROOM TABLE AT THE CUSTOMER SITE ON (B)(6) 2024 AND DETERMINED THE ACUTATOR COMPONENT WITHIN THE TABLE'S ROTATIONAL LOCKING MECHANISM CAUSED THE MALFUNCTION. THE SERVICE ENGINEER REMOVED AND REPLACED THE ACTUATOR COMPONENT AND FUNCTIONALLY TESTED THE TABLE AFTER THE REPAIR. INTERNAL CORRECTIVE ACTION IS IN PROCESS TO FURTHER ADDRESS THIS ISSUE.