inforMED
MalfunctionBWN

PIVOTPOST & CLAMP

Received Jan 15, 2025 · Event occurred Dec 19, 2024

Report 1220246-2025-09463 · MDR key 21157616

Device

Generic name

Operating Room Table Access

Manufacturer

Arthrex, Inc.

Model number

PIVOTPOST & CLAMP

Catalog number

AR-1560

Lot number

012434

Product problems

  • Break
  • Break

Patient

NA · Unknown

  • Insufficient Information
  • Insufficient Information

Narrative

Description of Event or Problem

ON 12/19/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS-06739377 THAT AN AR-1560 PIVOT POST AND CLAMP HAD THE PIECE THAT MAKES IT SWIVEL BREAK OFF. NO FURTHER INFORMATION WAS PROVIDED. THIS WAS DISCOVERED DURING A PROCEDURE. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

Additional Manufacturer Narrative

INVESTIGATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE PROVIDED UPON AVAILABILITY OF ADDITIONAL INFORMATION.

Additional Manufacturer Narrative

ADDITIONAL INFORMATION: D9, G3, H3, H6. COMPLAINT ALLEGATION IS CONFIRMED. ONE UNPACKAGED, AR-1560-02 & AR-1560-03. PIVOT POST AND CLAMP, SERIAL/BATCH NUMBER: (B)(6), WAS RECEIVED FOR INVESTIGATION. VISUAL EVALUATION NOTED THAT THE ADJUSTMENT KNOB WAS BROKEN OFF. FUNCTIONAL TESTING OF THE AR-1560-02 WAS NOT CONDUCTED DUE TO THE DAMAGE TO THE ROTATION LEVER. IT WAS NOTED THAT THE PIVOTPOST CLAMP WAS NOT DAMAGED AND FUNCTIONED AS INTENDED. THE MOST LIKELY CAUSE FOR THE REPORTED FAILURE CAN BE ATTRIBUTED TO MISUSE/MISHANDLING DUE TO THE DAMAGE TO THE DEVICE.