Received Jan 8, 2025 · Event occurred Dec 17, 2024

Report 2024168-2025-00292 · MDR key 21100847

Device

Generic name

Coronary Stent Delivery System

Model number

1012817-19

Catalog number

1012817-19

Lot number

3032341

Product problems

  • Failure to Advance
  • Failure to Advance

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

MANUFACTURER'S INVESTIGATION IS STILL PENDING AT THIS TIME. RESULTS AND CONCLUSIONS WILL BE PROVIDED IN THE FINAL REPORT.

Description of Event or Problem

IT WAS REPORTED THAT THE 2.8X19MM GRAFTMASTER COVERED STENT FAILED TO CROSS THE HEAVILY CALCIFIED AND HEAVILY TORTUOUS LESION IN THE LEFT ANTERIOR DESCENDING CORONARY ARTERY TO SEAL A PERFORATION. ANOTHER GRAFTMASTER COVERED STENT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY. NO ADDITIONAL INFORMATION WAS PROVIDED.

Additional Manufacturer Narrative

A VISUAL INSPECTION WAS PERFORMED ON THE RETURNED DEVICE. THE REPORTED FAILURE TO ADVANCE COULD NOT BE REPLICATED IN A TESTING ENVIRONMENT AS IT WAS RELATED TO OPERATIONAL CONTEXT OF THE PROCEDURE. PRODUCTION RECORD AND CORRECTIVE AND PREVENTATIVE ACTIONS (CAPA) REVIEWS WERE PERFORMED AND REVEALED NO INDICATION OF A PRODUCT QUALITY ISSUE. ADDITIONALLY, A QUERY OF THE COMPLAINT HANDLING DATABASE FOR THE REPORTED LOT REVEALED THERE IS NO INDICATION OF A LOT SPECIFIC ISSUE. BASED ON THE REPORTED INFORMATION AND THE OBSERVATIONS FROM THE RETURNED ANALYSIS, THE INVESTIGATION DETERMINED THAT THE REPORTED ISSUE APPEARS TO BE RELATED TO CIRCUMSTANCES OF THE PROCEDURE. BASED ON THE RESULTS OF THE COMPLAINT INVESTIGATION THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO THE DESIGN, MANUFACTURE, OR LABELING OF THE DEVICE.