MalfunctionJGS

VITROS CHEMISTRY PRODUCTS NA+ SLIDES

Received Jan 6, 2025 · Event occurred Dec 12, 2024

Report 1319809-2025-00001 · MDR key 21084735

Device

Generic name

In-vitro Diagnostics

Catalog number

8379034

Lot number

4260-1140-5329

Product problems

  • High Test Results
  • Non Reproducible Results
  • High Test Results
  • Non Reproducible Results

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE INVESTIGATION DETERMINED THAT A NON-REPRODUCIBLE, HIGHER THAN EXPECTED SODIUM (NA+) RESULT WAS OBTAINED FROM A SINGLE PATIENT SAMPLE PROCESSED USING VITROS CHEMISTRY PRODUCTS NA+ SLIDES LOT 4260-1140-5329 ON A VITROS XT 7600 INTEGRATED SYSTEM. THE INVESTIGATION COULD NOT DETERMINE A DEFINITIVE ASSIGNABLE CAUSE OF THE EVENT. BASED ON HISTORICAL QUALITY CONTROL RESULTS, A VITROS NA+ LOT 4260-1140-5329 PERFORMANCE ISSUE IS NOT A LIKELY CONTRIBUTOR OF THE EVENT. HOWEVER, AN INDIVIDUAL SLIDE RELATED ISSUE COULD NOT BE ENTIRELY RULED OUT. CONTINUAL TRACKING AND TRENDING OF COMPLAINTS HAS NOT IDENTIFIED ANY SIGNALS THAT WOULD POINT TO A POTENTIAL SYSTEMIC ISSUE WITH VITROS NA+ LOT 4260-1140-5329. AN INSTRUMENT RELATED ISSUE COULD NOT BE ENTIRELY RULED OUT AS A CONTRIBUTING FACTOR OF THE EVENT AS DIAGNOSTIC PRECISION TESTING WAS NOT PERFORMED ON THE VITROS XT 7600 INTEGRATED SYSTEM. HOWEVER, THERE WAS NO EVIDENCE INDICATING THAT THE INSTRUMENT MALFUNCTIONED. PRE-ANALYTICAL SAMPLE PROCESSING COULD NOT BE RULED OUT AS A CONTRIBUTING FACTOR AS THE CUSTOMER WAS NOT FOLLOWING THE SAMPLE COLLECTION DEVICE MANUFACTURER'S RECOMMENDED CENTRIFUGATION PROTOCOL. IMPROPER PRE-ANALYTICAL SAMPLE HANDLING COULD HAVE CONTRIBUTED TO THIS EVENT. IT IS POSSIBLE THAT CELLULAR DEBRIS, DUE TO POOR SAMPLE PREPARATION, WAS PRESENT IN THE AFFECTED SAMPLE, ALTHOUGH THIS COULD NOT BE CONFIRMED. THE CUSTOMER STATED THAT THEY SUSPENDED USE OF THE VITROS XT ALB-TP SLIDES ON (B)(6) 2024 AND THOROUGHLY CLEANED THE VITROS XT 7600 SYSTEM AS INSTRUCTED TO DO SO WITHIN A CUSTOMER LETTER SENT OUT BY ORTHO. THEREFORE, AN ISSUE RELATED TO THE USE OF VITROS XT ALB-TP SLIDES IS NOT A LIKELY CONTRIBUTOR OF THE NON-REPRODUCIBLE, HIGHER THAN EXPECTED VITROS NA+ RESULT.

Description of Event or Problem

A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT A NON-REPRODUCIBLE, HIGHER THAN EXPECTED SODIUM (NA+) RESULT WAS OBTAINED FROM A SINGLE PATIENT SAMPLE PROCESSED USING VITROS CHEMISTRY PRODUCTS NA+ SLIDES LOT 4260-1140-5329 ON A VITROS XT 7600 INTEGRATED SYSTEM. PATIENT SAMPLE RESULT OF >250 MMOL/L VS AN EXPECTED RESULT OF 125.5 MMOL/L BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED. THE NON-REPRODUCIBLE, HIGHER THAN EXPECTED PATIENT SAMPLE RESULT WAS NOT REPORTED OUTSIDE OF THE LABORATORY. THERE HAS BEEN NO REPORTED ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER (B)(4) AND REPORTABILITY ASSESSMENT (B)(4).