TINA-QUANT HEMOGLOBIN A1CDX GEN. 3

Received Dec 23, 2024 · Event occurred Dec 5, 2024

Report 1823260-2024-03801 · MDR key 21004197

Device

Generic name

Glycosylated Hemoglobin Assay

Manufacturer

Roche Diagnostics

Catalog number

08056668190

Lot number

ASKU

Product problems

  • Appropriate Device Problem Term/Code Not Available
  • Appropriate Device Problem Term/Code Not Available

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE SERIAL NUMBER OF THE C 303 ANALYZER IS (B)(6). THE INVESTIGATION IS ONGOING.

Description of Event or Problem

THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR ONE PATIENT SAMPLE TESTED WITH TINA-QUANT HEMOGLOBIN A1CDX GEN. 3 ON A COBAS PURE C 303 ANALYTICAL UNIT. THE SAMPLE INITIALLY RESULTED IN AN HBA1C VALUE OF 10.5 % AND IT REPEATED AS 5.8 %.

Additional Manufacturer Narrative

THE INVESTIGATION FINDINGS, INVESTIGATION CONCLUSION, AND COMPONENT CODES HAVE BEEN UPDATED.

Additional Manufacturer Narrative

THE CELL WASH WAS NOT OPTIMAL AND IT WAS IMPROVED AFTER ADJUSTMENTS PERFORMED BY THE FIELD SERVICE ENGINEER. THE INVESTIGATION DETERMINED THE ISSUE IS CONSISTENT WITH WATER DROPS IN THE REACTION CELLS. THE ROOT CAUSE OF WATER DROPLETS IN THE CELLS COULD NOT BE DETERMINED.

Additional Manufacturer Narrative

A HARDWARE CHECK INDICATED THERE WERE ISSUES WITH THE CELL RINSE ON THE ANALYZER. THE INVESTIGATION DETERMINED THE ISSUE IS CONSISTENT WITH AN ISSUE WITH CELL RINSE HARDWARE ADJUSTMENTS. THE CUSTOMER DID NOT REPORT ANY OTHER ISSUES.