inforMED
MalfunctionCEK

ATELLICA CH URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP)

Received Dec 19, 2024 · Event occurred Nov 25, 2024

Report 2432235-2024-00353 · MDR key 20979393

Device

Generic name

Total Protein

Model number

N/A

Catalog number

11097543

Lot number

140211

Product problems

  • Incorrect, Inadequate or Imprecise Result or Readings
  • Incorrect, Inadequate or Imprecise Result or Readings

Patient

49 YR · Female

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

ADDITIONAL INFORMATION (08-MAY-2025): SIEMENS SERVICE OPERATIONS SUPPORT (SOS) CONCLUDED THE INVESTIGATION OF THE EVENT. SOS REVIEWED THE INFORMATION PROVIDED BY THE CUSTOMER AND THE DATA FROM THE INSTRUMENT. NO REAGENT OR INSTRUMENT ISSUES WERE IDENTIFIED. QUALITY CONTROL (QC) RECOVERED WITHIN THE CUSTOMER¿S LABORATORY RANGES AT THE TIME OF THE EVENT. SIEMENS RECEIVED THE PATIENT SAMPLE FOR FURTHER TESTING AND INVESTIGATION. THE PATIENT SAMPLE WAS TESTED NEAT AND AUTO-DILUTED ON THE ATELLICA CH 930 ANALYZER. THE INTERNAL TESTING OF THE PATIENT SAMPLE AT SIEMENS OBTAINED THE SAME INCONSISTENT NEAT AND DILUTION RECOVERY THE CUSTOMER OBSERVED. BASED ON THE AVAILABLE INFORMATION, THE CAUSE OF THE EVENT IS SAMPLE SPECIFIC. A POTENTIAL PRODUCT ISSUE WAS NOT IDENTIFIED. THE CUSTOMER IS OPERATIONAL. THE DEVICE IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION IS REQUIRED. IN SECTION H6, THE INVESTIGATION FINDING AND INVESTIGATION CONCLUSION CODES WERE UPDATED. INITIAL MDR 2432235-2024-00353 WAS FILED ON 19-DEC-2024. MDR 2432235-2024-00352 WAS FILED FOR THE SAME EVENT.

Description of Event or Problem

THE CUSTOMER REPORTED TO SIEMENS THAT THEY OBTAINED DISCORDANT URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP) URINE PATIENT SAMPLE RESULTS ON AN ATELLICA CH 930 ANALYZER WHEN MANUALLY DILUTED. THE MANUAL DILUTED RESULTS WERE WITHIN THE MEASURING INTERVAL OF THE ASSAY AND WERE NOT REPORTED TO THE PHYSICIAN. THE SAME PATIENT SAMPLE WAS PREVIOUSLY PROCESSED ON THE ORIGINAL ATELLICA CH 930 ANALYZER NEAT AND AUTO DILUTED. THE NEAT RESULT WAS ABOVE ASSAY RANGE AND THE AUTO DILUTED RESULT WAS DEPRESSED. THE RESULTS WERE NOT REPORTED TO THE PHYSICIAN. THE CUSTOMER WAS NOT ABLE TO DETERMINE THE CORRECT RESULT FOR THE PATIENT. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP) RESULT.

Additional Manufacturer Narrative

AN OUTSIDE UNITED STATES (OUS) CUSTOMER CONTACTED THE SIEMENS HEALTHCARE DIAGNOSTICS REMOTE SERVICES CENTER (RSC) REGARDING DISCORDANT URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP) URINE PATIENT SAMPLE RESULTS ON AN ATELLICA CH 930 ANALYZER WHEN MANUALLY DILUTED. SIEMENS HEALTHCARE DIAGNOSTICS IS INVESTIGATING THE EVENT. A SEPARATE MDR WAS FILED FOR THE SAME EVENT.