WATCHMAN FLX DEVICE
Received Dec 18, 2024 · Event occurred Dec 5, 2024
Report MW5163714 · MDR key 20976833
Device
Generic name
System, Appendage Closure, Left Atrial
Patient
NA · Male
- Air Embolism
- Brain Injury
- Air Embolism
- Brain Injury
Narrative
Description of Event or Problem
PATIENT HAD A LEFT ATRIAL APPENDAGE OCCLUSION PROCEDURE WITH A WATCHMAN DEVICE. THERE WAS SYSTEMIC AIR EMBOLISM THAT CAUSED BRAIN INJURY AND ULTIMATELY DEATH OF THE PATIENT. AIR EMBOLISM WAS PARTLY DUE TO PATIENT TAKING DEEP BREATH AND SNORING WHILE UNDER MAC DURING THE PROCEDURE. THIS PROCEDURE IS DONE UNDER MAC IN SEVERAL INSTITUTIONS. HOWEVER, ABSENCE OF A HEMOSTATIC VALVE IN THE WATCHMAN SHEATH ALLOWED AIR TO ENTER THE SHEATH (WHEN INTRODUCING PIGTAIL CATHETER AND WATCHMAN FLX DEVICE) AND WAS SUCKED INTO THE LEFT ATRIUM AND THEREBY INTO THE SYSTEMIC CIRCULATION. PRESENCE OF A HEMOSTATIC VALVE WOULD HAVE PREVENTED THIS COMPLICATION.