ATLAS
Received Dec 19, 2024 · Event occurred Nov 19, 2024
Report 3005899764-2024-00075 · MDR key 20976597
Device
Generic name
Atlas Transfer Carriage
Manufacturer
Steris Mexico, S. De R.l. De C.v.Model number
AX0660230830000013Catalog number
AX0660230830000013
Product problems
- Failure to Align
- Failure to Align
Patient
NA · Unknown
- Muscle/Tendon Damage
- Muscle/Tendon Damage
Narrative
Additional Manufacturer Narrative
A STERIS SERVICE TECHNICIAN ARRIVED ON SITE TO INSPECT THE LOADING CAR ASSEMBLY AND FOUND THAT THE LOADING CAR WAS NOT PROPERLY ALIGNED WITH THE STERILIZER'S DOCKING STATION. TO RESOLVE THE ISSUE, THE TECHNICIAN ADJUSTED THE LOADING CAR, TESTED THE UNIT, CONFIRMED IT TO BE OPERATING ACCORDING TO SPECIFICATION, AND RETURNED IT TO SERVICE. NO ADDITIONAL ISSUES HAVE BEEN REPORTED.
Description of Event or Problem
THE USER FACILITY REPORTED THAT AN EMPLOYEE SUSTAINED A MUSCLE STRAIN IN THEIR SHOULDER WHILE UNLOADING THEIR LOADING CAR. MEDICAL TREATMENT WAS SOUGHT AND ADMINISTERED (FIRST AID).