inforMED
MalfunctionEXF

"D" BAG DRAINAGE SET

Received May 4, 2011 · Event occurred Apr 11, 2011

Report 1036710-2011-00005 · MDR key 2097432

Device

Generic name

Drainage Set

Model number

DBAG600

Catalog number

NA

Lot number

030823VI

Product problems

  • Device Operates Differently Than Expected

Patient

Not reported

  • No Information

Narrative

Description of Event or Problem

THE CUSTOMER REPORTED THE POUCH IS DRILLED.

Additional Manufacturer Narrative

THE PRODUCT WAS RETURNED FOR REVIEW. THE SEAL OF THE POUCH IS COMPLETE. THERE IS A STRAIGHT CUT MEASURING 3 INCHES ON THE POLY SIDE OF THE POUCH. THIS PRODUCT IS PACKAGED TWENTY TO A BOX. THE TWO FLAPS ON THE END OF THE BOX ARE FOLDED OVER AND TAPED. IF A CUTTING UTENSIL IS USED TO OPEN THE BOX AND IS PUSHED DEEP ENOUGH WHILE CUTTING THE TAPE, THE TOP POUCH MAY BE CUT AS SEEN ON THE RETURNED POUCH. THIS IS THE FIRST REPORT OF THIS OCCURRING. THE DEFECT IS READILY VISIBLE BY THE CUSTOMER. WE WILL CONTINUE TO MONITOR AND TREND TO TAKE ACTION WHEN NECESSARY.