PORTEX CUFF PRESSURE MONITORING

Received Dec 17, 2024 · Event occurred Jul 1, 2024

Report 3012307300-2024-14170 · MDR key 20956697

Device

Generic name

Monitor, Airway Pressure (includes Gauge And/or Alarm)

Catalog number

54-05-000JJP

Lot number

UNKNOWN

Product problems

  • Gas/Air Leak
  • Gas/Air Leak

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

B3. DATE OF EVENT: UNKNOWN. NO INFORMATION HAS BEEN PROVIDED TO DATE. DEVICE EVALUATION: ONE DEVICE WAS RECEIVED. A VISUAL INSPECTION FOUND NO ABNORMALITIES. DURING THE FUNCTIONAL TESTING, THERE WERE NO ABNORMALITIES FOUND IN THE PRODUCT AND IT WORKED NORMALLY. WHEN THE TUBE THAT CAME WITH THE PRODUCT WAS CONNECTED TO THE MAIN UNIT, THE REPORTED PHENOMENON WAS REPRODUCED. THE CAUSE IS AN AIR LEAK IN THE ATTACHED TUBE. THE CUFF PRESSURE GAUGE DEVICE PASSED ALL FUNCTIONAL TESTS.

Description of Event or Problem

IT WAS REPORTED THAT "THE NEEDLE COMES DOWN." THERE WAS PATIENT INVOLVEMENT. PER INVESTIGATION: DURING THE FUNCTIONAL TESTING, THERE WERE NO ABNORMALITIES FOUND IN THE PRODUCT, AND IT WORKED NORMALLY. WHEN THE TUBE THAT CAME WITH THE PRODUCT WAS CONNECTED TO THE MAIN UNIT, THE REPORTED PHENOMENON WAS REPRODUCED. THE CAUSE WAS AN AIR LEAK IN THE ATTACHED TUBE.