RBI2 SUCTION RECTAL BIOPSY SYSTEM

Received Dec 13, 2024 · Event occurred Oct 23, 2024

Report 3005921952-2024-00004 · MDR key 20923370

Device

Generic name

Rbi2

Model number

RBI2

Catalog number

CP1200

Lot number

22409

Product problems

  • No Apparent Adverse Event
  • No Apparent Adverse Event

Patient

4 MO · Male

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

COMPARISON OF BLADE SIMULATION IN THE SAME POSITION AS RETURNED TIP ONLY. POSITION MATCHES RETURNED SETTING, SHOWING CONTACT WITH THE CUTTING BOARD BUT CUTTING BOARD NOT FULLY CUT. THE RETURNED CAPSULE SHOWS EVIDENCE OF JAMMED TISSUE AT THE POINT OF THE CUTTING BOARD AND THE MIDDLE OF THE BLADE. THE BLADE WAS NOT FULLY ACTIVATED DURING INITIAL USE AND STOPPED AT THE FIRST RESISTANCE. UPON MANUAL ACTIVATION, THE BLADE FIRED CORRECTLY INTO ITS FINAL POSITION, AND THE CAPSULE SNAPPED AS EXPECTED. PICTURES CONFIRMED THE BLADE, ONCE FULLY ACTIVATED, RELEASED THE JAMMED TISSUE. SIMULATIONS SHOWED PROPER FUNCTIONING OF THE BLADE AND CUTTING BOARD WHEN FULLY ACTIVATED, RULING OUT MANUFACTURING FAULTS. THIS JAMMING PREVENTED PROPER TISSUE ACTIVATION BY THE CUTTING MECHANISM. ISSUE LINKED TO INCOMPLETE BLADE ACTIVATION AND IMPROPER INTERACTION WITH THE CUTTING BOARD. THIS CAUSED A FAILURE TO ACHIEVE A CLEAN CUT AND RELEASE OF THE SPECIMEN THE ROOT CAUSE OF THE ISSUE WAS THE BLADE FAILING TO FULLY ACTIVATE DURING USE, LEADING TO TISSUE JAMMING. THIS FINDING IS CONCLUSIVE AND BEYOND DOUBT, WITH NO ADDITIONAL DEFECTS NOTED IN THE CAPSULE OR CUTTING MECHANISM UPON FURTHER ANALYSIS.

Description of Event or Problem

"WE DID CONFIRM WITH A DRY RUN THAT THE BLADE WOULD DEPLOY WHEN FIRED, HOWEVER WHEN ATTEMPTING TO GET A SPECIMEN EVEN AFTER THIS WE WERE UNABLE TO GET A BIOPSY SPECIMEN. WE SWITCHED THE GUN TO SUCTION RECTAL BIOPSY GUN # 4 AND STILL HAD THE SAME ISSUE UNTIL ABOUT THE 3RD OR 4TH ATTEMPT WHEN THE GUN FINALLY WAS ABLE TO GET 2 SPECIMENS. IN EACH ATTEMPT AT BIOPSY, THE TRIGGER WAS PUSHED THE ENTIRE WAY TO THE METAL END SUCH THAT IT COULD NOT HAVE BEEN DEPLOYED FURTHER."