CADD-SOLIS COMM MODULE
Received Dec 12, 2024 · Event occurred May 21, 2024
Report 3012307300-2024-14825 · MDR key 20920025
Device
Generic name
Accessories, Pump, Infusion
Manufacturer
Smiths Medical Asd, Inc.Catalog number
21-2131-01
Lot number
18766925
Product problems
- Wireless Communication Problem
- Wireless Communication Problem
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
ONE DEVICE WAS RECEIVED WITH THE HOST CADD SOLIS PUMP TO BE EVALUATED TOGETHER. THE EVALUATION TEST FOUND THE FAILURE OCCURRING WHEN TESTED ON ITS HOST PUMP WITH THE POLE MOUNT CAPTURE MOUNTED BUT NOT REPEATING WHEN TESTED ON A KNOWN GOOD TEST SOLIS PUMP. DETERMINED THAT THE FAULT WAS WITH THE HOST PUMP NOT THE COMM MODULE. CORRECTIVE ACTIONS ARE IN PROCESS.
Description of Event or Problem
IT WAS REPORTED THAT UPON IMPLEMENTATION, THE PUMP HAD AN OUT OF BOX FAILURE WITH WIRELESS INTERMITTENT CONNECTION WITH PUMP. THE EVENT OCCURRED DURING IMPLEMENTATION OF PUMP AT FACILITY. THERE WAS NO PATIENT INVOLVEMENT.