inforMED
MalfunctionMRZ

CADD-SOLIS COMM MODULE

Received Dec 12, 2024 · Event occurred May 21, 2024

Report 3012307300-2024-14825 · MDR key 20920025

Device

Generic name

Accessories, Pump, Infusion

Catalog number

21-2131-01

Lot number

18766925

Product problems

  • Wireless Communication Problem
  • Wireless Communication Problem

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

ONE DEVICE WAS RECEIVED WITH THE HOST CADD SOLIS PUMP TO BE EVALUATED TOGETHER. THE EVALUATION TEST FOUND THE FAILURE OCCURRING WHEN TESTED ON ITS HOST PUMP WITH THE POLE MOUNT CAPTURE MOUNTED BUT NOT REPEATING WHEN TESTED ON A KNOWN GOOD TEST SOLIS PUMP. DETERMINED THAT THE FAULT WAS WITH THE HOST PUMP NOT THE COMM MODULE. CORRECTIVE ACTIONS ARE IN PROCESS.

Description of Event or Problem

IT WAS REPORTED THAT UPON IMPLEMENTATION, THE PUMP HAD AN OUT OF BOX FAILURE WITH WIRELESS INTERMITTENT CONNECTION WITH PUMP. THE EVENT OCCURRED DURING IMPLEMENTATION OF PUMP AT FACILITY. THERE WAS NO PATIENT INVOLVEMENT.