IDEAL SUTURE SHUTTLE 90 DEGREES
Received Dec 11, 2024 · Event occurred Nov 28, 2024
Report 1221934-2024-04423 · MDR key 20909594
Device
Generic name
Instrument, Ligature Passing And Knot Tying
Manufacturer
Depuy Mitek Llc UsCatalog number
251005
Lot number
24D04
Product problems
- Break
- Break
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THIS REPORT IS BEING SUBMITTED IN PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY MITEK OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MITEK, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. H11 ADDITIONAL NARRATIVE: D10. CONCOMITANT MED PRODUCTS AND THERAPY DATES: DYNATAPE DEVICE (B)(6) 2024. INVESTIGATION SUMMARY ==> THE COMPLAINT DEVICE WAS DISCARDED BY THE CUSTOMER, THEREFORE NOT RETURNED TO J&J MEDTECH ORTHOPAEDICS AND UNAVAILABLE FOR A PHYSICAL EVALUATION. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE LOT NUMBER, AND NO NON-CONFORMANCE WAS IDENTIFIED. BASED ON THE CURRENT AVAILABLE INFORMATION, WE CANNOT DETERMINE A ROOT CAUSE FOR THE REPORTED FAILURE. IF ADDITIONAL INFORMATION IS RECEIVED IN THE FUTURE, WE WILL REOPEN THE COMPLAINT AND PERFORM THE INVESTIGATION AS APPROPRIATE. IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED AND THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE REPORTED COMPLAINT CONDITION. AS PART OF J&J MEDTECH ORTHOPAEDICS QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES.
Description of Event or Problem
IT WAS REPORTED FROM JAPAN THAT DURING AN ARTHROSCOPIC ROTATOR CUFF REPAIR (ARCR) FOR A RIGHT SHOULDER TEAR, THE PHYSICIAN PASSED THE DYNATAPE THROUGH THE WIRE KITE AT THE TIP OF THE IDEAL SUTURE SHUTTLE DEVICE, INSIDE THE PATIENT'S JOINT, AND WHEN THE DYNATAPE WAS TAKEN OUT OF THE JOINT, THE WIRE BROKE. POSTOPERATIVE X-RAYS CONFIRMED THAT THERE WERE NO FRAGMENTS IN THE PATIENT'S BODY. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO SURGICAL DELAY AND NO HARM TO THE PATIENT.