IMMULITE® 2000 3GALLERGY SPECIFIC IGE

Received Dec 11, 2024 · Event occurred Nov 25, 2024

Report 1219913-2024-00496 · MDR key 20905172

Device

Generic name

System, Test, Radioallergosorbent (rast) Immunological

Model number

N/A

Catalog number

10380875

Lot number

145

Product problems

  • Low Test Results
  • Low Test Results

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

AN OUTSIDE UNITED STATES (OUS) CUSTOMER CONTACTED A SIEMENS CUSTOMER CARE CENTER TO REPORT DISCORDANT LOW 3GALLERGY¿ SPECIFIC IGE UNIVERSAL KIT RESULTS THAT WERE OBTAINED ON PATIENT SAMPLES ON AN IMMULITE 2000 XPI INSTRUMENT. QUALITY CONTROL (QC) LEVEL 1 WAS OBSERVED OUT OF RANGE. QC LEVEL 2 WAS IN RANGE. ADDITIONALLY, THE ADJUSTMENT HAD AN INTERCEPT HIGHER THAN THE GUIDE. THE LIMITATIONS SECTION OF THE IMMULITE 2000 3GALLERGY¿ SPECIFIC IGE UNIVERSAL KIT INSTRUCTIONS FOR USE (IFU) STATES: "FOR DIAGNOSTIC PURPOSES, THE RESULTS OBTAINED FROM THIS ASSAY SHOULD ALWAYS BE USED IN COMBINATION WITH THE CLINICAL EXAMINATION, PATIENT MEDICAL HISTORY, AND OTHER FINDINGS." SIEMENS IS INVESTIGATING. NOTE: THIS OUS PRODUCT (CATALOG NUMBER: 10380875, AS LISTED IN D4) IS ASSOCIATED WITH SIMILAR PRODUCT IN THE UNITED STATES: CATALOG NUMBER: 10708840 / PREMARKET APPROVAL (PMA)# K101572.

Description of Event or Problem

THE CUSTOMER REPORTED THE OBSERVATION OF DISCORDANT LOW 3GALLERGY¿ SPECIFIC IGE UNIVERSAL KIT RESULTS OBTAINED ON PATIENT SAMPLES ON AN IMMULITE 2000 XPI INSTRUMENT. THE INITIAL RESULTS WERE NOT REPORTED TO THE PHYSICIAN(S). THE SAMPLES WERE REPEATED ON THE SAME INSTRUMENT. THE REPEAT RESULTS WERE HIGHER THAN THE INITIAL RESULTS. IT IS UNKNOWN IF THE REPEAT RESULTS WERE REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THIS ISSUE.

Additional Manufacturer Narrative

SIEMENS FILED THE INITIAL MDR 1219913-2024-00496 ON 11-DEC-2024. ADDITIONAL INFORMATION 14-FEB-2025: SIEMENS EVALUATED THIS ISSUE AND PROVIDED THE FOLLOWING CONCLUSION: THE CUSTOMER REPORTED THE OBSERVATION OF AN INTERCEPT OUTSIDE OF CRITERIA, QC LEVEL 1 HIGH OUT OF RANGE, AND 3 PATIENTS INITIALLY RESULTED AS <0.1 KU/L AND ARE HIGHER ON REPEAT WITH IMMULITE 2000 3GALLERGY SPECIFIC IGE (SPE) KIT LOT 145. SIEMENS REVIEWED INTERNAL BOX AND WHISKER (B&W) DATA FOR QUALITY CONTROLS. THE B&W DATA IS WITHIN SIEMENS INSTRUCTIONS FOR USE (IFU) RANGE. TO EXCLUDE AN ISSUE WITH THE INSTRUMENT REGARDING THE DISCORDANT RESULTS, INSTRUMENT FILES WERE REQUESTED, BUT NOT RECEIVED. FURTHER EVALUATION REGARDING THE DISCORDANT RESULTS WAS NOT POSSIBLE. THE CUSTOMER USED FRESH SPE KIT LOT 145 AND FRESH CONTROLS AND INTERCEPTS AND QC ARE IN RANGE AND THE CUSTOMER IS NOW SATISFIED WITH ASSAY PERFORMANCE. THE ISSUE IS RESOLVED WITH ROUTINE TROUBLESHOOTING USING FRESH REAGENTS AND FRESH QC MATERIAL. THE CUSTOMER IS OPERATIONAL AND SATISFIED WITH THE PERFORMANCE. THE IMMULITE 2000 3GALLERGY SPECIFIC IGE (SPE) KIT LOT 145 IS PERFORMING AS INTENDED. A PRODUCT PERFORMANCE ISSUE HAS NOT BEEN IDENTIFIED. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED. IN SECTION H6, THE INVESTIGATION FINDINGS AND INVESTIGATION CONCLUSIONS CODES WERE UPDATED.