OSTEOTOME 6MM X 1.5 IN
Received Dec 10, 2024 · Event occurred Nov 1, 2024
Report 1818910-2024-25110 · MDR key 20888837
Device
Generic name
Hip Instruments : Cutting Instruments
Manufacturer
Depuy Orthopaedics Inc UsCatalog number
242241000
Lot number
VZ17695
Product problems
- Scratched Material
- Scratched Material
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THE DEVICE MALFUNCTION, VISUAL : SCRATCHED/NICKED, WILL NO LONGER BE REPORTED AS THERE HAS NOT BEEN A SERIOUS INJURY ASSOCIATED WITH THIS FAILURE WITHIN THE PAST 2 YEARS. PLEASE CONSIDER THIS THE LAST MANUFACTURER REPORT NUMBER SUBMITTED AGAINST THIS MALFUNCTION.
Additional Manufacturer Narrative
PRODUCT COMPLAINT # (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803 (AND/OR PART 4, AS APPLICABLE). THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. ADDITIONAL NARRATIVE: D4: UDI: AS THE LOT NUMBER FOR THE DEVICE INVOLVED IN THE EVENT WAS NOT PROVIDED, THE FULL UDI IS CURRENTLY NOT AVAILABLE. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.
Description of Event or Problem
IT WAS REPORTED THAT THE OSTEOTOMES WERE COMPLETELY DAMAGED BEFORE THE OPERATION. NO FURTHER INFORMATION IS PROVIDED.
Additional Manufacturer Narrative
PRODUCT COMPLAINT # (B)(4). INVESTIGATION SUMMARY : ACCORDING TO THE INFORMATION RECEIVED, "THE OSTEOTOMES WERE COMPLETELY DAMAGED BEFORE THE OPERATION." THE PRODUCT WAS NOT RETURNED TO MEDTECH ORTHOPAEDICS, HOWEVER PHOTOS WERE PROVIDED FOR REVIEW. THE PHOTO INVESTIGATION REVEALED THAT '242241000, OSTEOTOME 6MM X 1.5 IN' HAD SCRATCHED. POTENTIAL CAUSE CAN BE ATTRIBUTED TO UNINTENDED USE ERROR BY USE OF EXCESSIVE FORCE IN A PRYING MOTION AND OVERLOAD OF THE MATERIAL. SINCE THE DEVICE WAS NOT RETURNED, A DIMENSIONAL INSPECTION CANNOT BE PERFORMED. THE OVERALL COMPLAINT WAS CONFIRMED AS THE OBSERVED CONDITION OF THE OSTEOTOME 6MM X 1.5 IN WOULD CONTRIBUTE TO THE COMPLAINED DEVICE ISSUE. BASED ON THE INVESTIGATION FINDINGS, UNINTENDED USE ERROR CAUSED OR CONTRIBUTED TO EVENT AND IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE COMPLAINT CONDITION. AS PART OF MEDTECH ORTHOPAEDICS QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE HISTORY LOT : THE DEVICE LOT NUMBER IS UNKNOWN, THEREFORE A¿DEVICE HISTORY REVIEW COULD NOT BE PERFORMED. IF THE LOT/SERIAL NUMBER BECOMES AVAILABLE, THE RECORD WILL BE RE-ASSESSED.