InjuryFLL
S-CATH M ESOPHAGEAL TEMPERATURE PROBE
Received Dec 6, 2024 · Event occurred Nov 6, 2024
Report 3009437315-2024-00001 · MDR key 20869451
Device
Generic name
Esophageal Temperature Probe
Manufacturer
Circa Scientific, Inc.Model number
CS-46EPLot number
004814
Product problems
- Material Deformation
- Material Deformation
Patient
NA · Unknown
- Epistaxis
- Epistaxis
Narrative
Additional Manufacturer Narrative
EXAMINATION SHOWS THE PEBAX LAYER WAS THIN AT THE ELECTRODE BAND EDGES. DURING PROBE INSERTION AND REMOVAL, THE FRICTION CAUSED THE PEBAX TO PEEL AWAY, THEREBY EXPOSING THE ELECTRODE BAND EDGE.
Description of Event or Problem
THE CUSTOMER USED A NEW CS-46EP AND REPORTS THAT WHEN THE PROBE WAS REMOVED, THERE WAS RESISTANCE IN THE NOSE, AND IT STARTED TO BLEED HEAVILY; THE BLEEDING COULD ONLY BE STOPPED THE NEXT DAY. HE ALSO REPORTED THAT THE SURFACE OF THE PROBE IN THE AREA OF THE ELECTRODES WAS BADLY DAMAGED.