MalfunctionDRG
CERNER CAREAWARE IBUS
Received Dec 6, 2024 · Event occurred Jun 17, 2024
Report 20863586 · MDR key 20863586
Device
Generic name
Transmitters And Receivers, Physiological Signal, Radiofrequency
Manufacturer
Cerner CorporationModel number
CAREAWAREIBUS5Catalog number
CAREAWAREIBUS5
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
DURING AN UPGRADE TO OUR CERNER SYSTEMS, A MISTAKE WAS MADE IN A PORTION OF THE SOFTWARE THAT COMMUNICATED WITH OUR BAXTER INTRAVENOUS LINE (IV) SERVER THAT ALLOWED ERRORS TO BE PUT INTO THE PROGRAMING OF OUR IV PUMPS. THE ERROR WAS ONLY FOUND AFTER STAFF NOTICED THE PUMP THEY WERE PROGRAMING WASN'T INPUTTING THE VALUES THEY WERE ADDING TO IT SEVERAL DAYS AFTER THE UPGRADE.