CADD-SOLIS COMM MODULE
Received Dec 4, 2024 · Event occurred May 21, 2024
Report 3012307300-2024-14419 · MDR key 20849909
Device
Generic name
Accessories, Pump, Infusion
Manufacturer
Smiths Medical Asd, Inc.Catalog number
21-2131-01
Lot number
18766925
Product problems
- Wireless Communication Problem
- Wireless Communication Problem
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
IT WAS REPORTED, THAT UPON IMPLEMENTATION, THE PUMP HAD AN OUT OF BOX FAILURE. WITH WIRELESS INTERMITTENT CONNECTION WITH PUMP. THERE WAS NO PATIENT INVOLVEMENT.
Additional Manufacturer Narrative
ONE DEVICE WAS RECEIVED, WITH THE HOST CADD SOLIS PUMP TO BE EVALUATED TOGETHER. EVALUATION FOUND THE FAILURE OCCURRING, WHEN THE COMM MODULE IS TESTED ON ITS HOST PUMP, WITH AC POWER ADAPTER PLUGGED IN. AND IDENTIFIED, THE COMM MODULE WAS NOT FULLY LATCHING INSIDE ITS HOST PUMP. THE FAILURE OCCURRED, WITH OR WITHOUT A POLE MOUNT CAPTURE MOUNTED, BUT NOT REPEATING, WHEN A KNOWN GOOD TEST SOLIS PUMP WAS SWAPPED. DETERMINED, THE HOST PUMP WAS CONTRIBUTING TO THE FAILURE.