inforMED
MalfunctionCZP

TINA-QUANT IGA GEN.2

Received Dec 4, 2024 · Event occurred Nov 14, 2024

Report 1823260-2024-03514 · MDR key 20843740

Device

Generic name

Immunoglobulin A

Manufacturer

Roche Diagnostics

Catalog number

08057885190

Lot number

ASKU

Product problems

  • Low Test Results
  • Non Reproducible Results
  • Low Test Results
  • Non Reproducible Results

Patient

12 MO · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

NO QUESTIONABLE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. ALL RESULTS WERE ACCOMPANIED BY DATA FLAGS SINCE SYSTEM REAGENTS ON BOARD THE ANALYZER WERE EXPIRED. THE INVESTIGATION DETERMINED THE ISSUE WAS DUE TO A USER ERROR SINCE SYSTEM REAGENTS EXCEEDED ONBOARD STABILITY AT THE TIME OF THE EVENT.

Description of Event or Problem

THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR ONE PATIENT SAMPLE TESTED WITH TINA-QUANT IGA GEN.2 ON A COBAS PURE C 303 ANALYTICAL UNIT. THE SAMPLE INITIALLY RESULTED IN AN IGA VALUE OF 2.0 MG/DL. THE SAMPLE WAS REPEATED ON A SECOND ANALYZER, RESULTING IN A VALUE OF 29.04 MG/DL.

Additional Manufacturer Narrative

THE SERIAL NUMBER OF THE C 303 ANALYZER IS (B)(6). THE INVESTIGATION IS ONGOING.