I-STAT CREATININE CARTRIDGE
Received Dec 3, 2024 · Event occurred Nov 22, 2024
Report 2245578-2024-00209 · MDR key 20830797
Device
Generic name
Crea Cartridge
Manufacturer
Abbott Point Of Care Inc.Model number
NACatalog number
03P84-25
Lot number
A24156
Product problems
- Incorrect, Inadequate or Imprecise Result or Readings
- Incorrect, Inadequate or Imprecise Result or Readings
Patient
50 YR · Male
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
APOC INCIDENT: # (B)(4). THE INVESTIGATION WAS COMPLETED ON 16-DEC-2024. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) CONFIRMED THE CARTRIDGE LOT MET FINISHED GOODS (FG) RELEASE CRITERIA. RETAINED CARTRIDGE TESTING MET THE ACCEPTANCE CRITERIA OUTLINED IN APPENDIX 1 OF Q04.01.003 REV. AN, PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE. NO DEFICIENCY HAS BEEN IDENTIFIED FOR CREA CARTRIDGE LOT A24156.
Description of Event or Problem
ON 25-NOV-2024, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT CREA CARTRIDGES THAT YIELDED A SUSPECTED DISCREPANT CREATININE RESULTS ON A 50-YEAR-OLD MALE PATIENT WITH A NON-HEALING SURGICAL WOUND. THERE WAS NO ADDITIONAL PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT. METHOD: DATE: COLLECTED: TESTED: RESULT: (MG/DL) SAMPLE: I-STAT, ON (B)(6) 2024, 0952, 0954, >2.0, WB; I-STAT, NI, NI, NI, 1.5, NI; BECKMAN, NI, NI, NI, 1.46, NI. THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE INVESTIGATION IS UNDERWAY.
Additional Manufacturer Narrative
APOC INCIDENT#: (B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.