inforMED
MalfunctionCBF

PAHI HBO CHAMBER

Received Apr 18, 2011 · Event occurred Dec 1, 2010

Report 2075477 · MDR key 2075477

Device

Generic name

Hyperbaric Oxygen Chamber

Model number

*

Catalog number

*

Lot number

*

Product problems

  • Decrease in Pressure
  • Device Dislodged or Dislocated

Patient

65 YR

  • Pain
  • Dizziness

Narrative

Description of Event or Problem

PT WAS TREATED FOR DIABETIC FOOT ULCER. PT'S 25TH TREATMENT WITH HISTORY OF SINUS CONGESTION. THE DOOR LATCH DISLODGED, RESULTING IN SUDDEN DECOMPRESSION OF THE CHAMBER. PT C/O COMPLAINED OF LIGHT HEADEDNESS AND EAR TENDERNESS. STAFF NOTIFIED VENDOR ABOUT THE INCIDENT AND PAN-AMERICA HYPERBARICS INC RESPONDED WITH A SITE VISIT WITHIN 24 HOURS. PAHI REPLACED THE OUTER STEEL LATCH HANDLE AND ULTIMATELY REPLACED THE ENTIRE DOOR LATCH OF THIS CHAMBER AND THE OTHER HYPERBARIC OXYGEN CHAMBER.======================MANUFACTURER RESPONSE FOR HYPERBARIC OXYGEN CHAMBER, PAHI HBO CHAMBER======================MANUFACTURER FIXED THE DOOR.