COUPLER AC FFL 19
Received Nov 20, 2024 · Event occurred Jan 25, 2024
Report 1221934-2024-04154 · MDR key 20737200
Device
Generic name
Accessories, Photographic, For Endoscope (exclude Light Sources)
Manufacturer
Medos International Sã rlCatalog number
242437
Product problems
- Break
- Unintended Movement
- Break
- Unintended Movement
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE DEVICE WAS EXTREMELY LOOSE. DURING THE SERVICE EVALUATION THE FOLLOWING DEFECTS WERE IDENTIFIED: BROKEN (2+ PIECES): CHIPPED/SHAVED/DELAMINATED. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH AN UNSPECIFIED DELAY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Additional Manufacturer Narrative
THIS REPORT IS BEING SUBMITTED IN PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY MITEK OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY MITEK, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. H11 ADDITIONAL NARRATIVE: H4: DEVICE MANUFACTURE DATE IS CURRENTLY UNAVAILABLE. INVESTIGATION SUMMARY THE COMPLAINT DEVICE WAS RECEIVED AT THE SERVICE CENTER AND EVALUATED. DURING THE SERVICE EVALUATION THE FOLLOWING DEFECTS WERE IDENTIFIED: BROKEN (2+ PIECES) : CHIPPED/SHAVED/DELAMINATED, DEVICE INTERACTION (2+ DEVICES) : WILL NOT HOLD/RETAIN. - DAMAGED GRASPING MECHANISM - EXTERNAL DAMAGE OF ADJUSTING RING CAUSE OF FAULT: - QUICK COUPLER WORN OUT - DAMAGE CAUSED BY INCORRECT HANDLING PER SERVICE REPORTS, THIS COMPLAINT WAS CONFIRMED. BASED ON THE INVESTIGATION FINDINGS, THE USER ERROR WERE IDENTIFIED AS THE ROOT CAUSE FOR THE DEVICE FAILURE DURING THE SERVICE EVALUATION. IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED AND THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE REPORTED COMPLAINT CONDITION. AS PART OF J&J MEDTECH ORTHOPAEDICS QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES.
Additional Manufacturer Narrative
THIS REPORT IS BEING SUBMITTED IN PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY MITEK OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY MITEK, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. INVESTIGATION SUMMARY: CORRECTED THE COMPLAINT DEVICE WAS RECEIVED AT THE SERVICE CENTER AND EVALUATED. DURING THE SERVICE EVALUATION THE FOLLOWING DEFECTS WERE IDENTIFIED: DAMAGED GRASPING MECHANISM, EXTERNAL DAMAGE OF ADJUSTING RING BROKEN (2+ PIECES) : CHIPPED/SHAVED/DELAMINATED, AND DEVICE INTERACTION (2+ DEVICES) : WILL NOT HOLD/RETAIN PER SERVICE REPORT, THIS COMPLAINT WAS CONFIRMED. BASED ON THE INVESTIGATION FINDINGS, THE POTENTIAL ROOT CAUSE IS TRACED TO USER ERROR. IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED AND THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE REPORTED COMPLAINT CONDITION. AS PART OF J&J MEDTECH ORTHOPAEDICS QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES.