inforMED
InjuryEGJ

HYBRESIS

Received Apr 29, 2011 · Event occurred Mar 23, 2011

Report 1721293-2011-00002 · MDR key 2073089

Device

Generic name

Hybresis

Manufacturer

Empi, Inc.

Model number

199589-001

Lot number

83670

Product problems

  • Insufficient Device Problem Information

Patient

49 YR

  • Pain
  • Full thickness (Third Degree) Burn

Narrative

Additional Manufacturer Narrative

THE ACTUAL DEVICE WAS NOT RETURNED FOR EVALUATION. THREE UNUSED PATCHES OF THE SAME LOT NUMBER WERE RETURNED. ALL MET PHYSICAL AND ELECTRICAL SPECIFICATIONS. THE USER MANUAL FOR THE HYBRESIS PATCH STATES: ADVISE THE PATIENT TO REMOVE ELECTRODES IF ANY UNDUE SENSATION OF PAIN OR BURNING OCCURS DURING THE TREATMENT AND TO REPORT DISCOMFORT TO CLINIC. APPLY THE HYDRATED PATCH SO THAT THE DRUG PAD IS OVER THE TREATMENT SITE AND SECURE IT BY PRESSING THE ADHESIVE BORDER. AVOID PRESSING DIRECTLY OVER THE PADS. PRESSING DIRECTLY ON THE PADS CAN CAUSE LEAKAGE THAT WILL COMPROMISE ADHESION TO THE PATIENT. NOTE: DO NOT TAPE OR BIND THE PATCH DURING TREATMENT. DO NOT APPLY HOT OR COLD THERAPY OVER PATCH DURING TREATMENT. THE DEVICE HISTORY RECORD WAS REVIEWED, AND THE LOT OF PATCHES MET ALL RELEASED SPECIFICATIONS. THIS INCIDENT IS SIMILAR TO A COMPLAINT RECEIVED BY EMPI THAT SUGGESTS THAT THIS DEVICE MAY HAVE MALFUNCTIONED IN SUCH A WAY THAT IT COULD HAVE CONTRIBUTED TO THIS INJURY.

Description of Event or Problem

IT WAS REPORTED THAT A PATIENT RECEIVED A 3RD DEGREE BURN ON TOP OF HER RIGHT FOOT WHILE USING AN EMPI HYBRESIS DEVICE; THIS WAS THE 4TH TREATMENT. THE COMPOUND USED WAS DEXAMETHASONE AT 4MG/ML, AND 1.5 ML WAS USED ON THE PATCH. THE MEDICATION WAS PLACED ON THE NEGATIVE SIDE AND SALINE WAS PLACED ON THE POSITIVE SIDE OF THE PATCH. THE MODE USED WAS "HYBRESIS", AND THE DOSE CONTROLLER WAS ON FOR 3 MINUTES, AND THE PATCH WAS WORN FOR 2 HOURS. BEFORE THE TREATMENT THE PATIENT HAD AN ULTRASOUND, AND MYOFASCIAL RELEASE OF THE LOWER LEG. A COBAN WRAP WAS USED TO WRAP OVER THE ELECTRODE, AND THE PATIENT PUT A TENNIS SHOE ON. THE PATIENT REPORTED HAVING SEVERE PAIN AND TRIED TO LEAVE THE PATCH ON AND TOLERATE THE PAIN. WHEN THE PATCH WAS REMOVED THE PATIENT NOTICED A BURN UNDER THE PATCH AND RETURNED TO THE CLINIC. SHE WAS INSTRUCTED TO FOLLOW UP WITH HER MD; SHE SAW HER MD THE NEXT DAY. THE BURN WAS NOTED AS BEING 3RD DEGREE, AND BY THE PICTURES SENT THE BURN LOOKS TO BE 1CM IN DIAMETER. WOUND WAS DEBRIDED.