STERNAL ZIPFIX CABLE TIE W/NEEDLE PEEK 2
Received Nov 15, 2024 · Event occurred Nov 7, 2024
Report 8030965-2024-13990 · MDR key 20695985
Device
Product problems
- Break
- Break
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
Description of Event or Problem
IT WAS REPORTED THAT ON NOVEMBER 7, 2024, THE EVENT INVOLVED 3 DEVICES. THE FIRST AND SECOND DEVICES COULD NOT TIGHTEN THE ZIPFIX. THE THIRD DEVICE (ZIPFIX) HAD BREAKAGE. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Additional Manufacturer Narrative
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H3, H6: THE PRODUCT WAS RETURNED TO J&J MEDTECH ORTHOPAEDICS FOR EVALUATION. VISUAL INSPECTION OF THE RETURNED SAMPLE FOUND THE IMPLANT WAS SEPARATED IN THREE FRAGMENTS. ONE FROM THE NEEDLE NOTCH AND THE SECOND FROM THE TOOTH ATTACHING SURFACE. THEREFORE, THE REPORTED CONDITION CAN BE CONFIRMED. NO OTHER STRUCTURAL ANOMALIES WERE IDENTIFIED. A COMPLETE POTENTIAL CAUSE FOR THE OBSERVED CONDITION CAN NOT BE DETERMINED WITH AVAILABLE INFORMATION. PROPERLY HANDLING AND ATTENTION TO THE APPROVED USE OF THE DEVICE DIMINISHES THE RISK OF FAILURE. SURGICAL TECHNIQUE SE_839363 REV. AB WAS REVIEWED AND THE FOLLOWING RELEVANT STATEMENTS WERE FOUND AT PAGE 10: PRECAUTIONS: TO AVOID DAMAGE TO THE LOCKING HEAD: PRIOR TO INSERTION OF THE CUT END, ENSURE THE ZIPFIX IMPLANT IS PROPERLY ORIENTED SUCH THAT THE TOOTHED SURFACE CONTACTS THE STERNUM. ALIGN THE CUT END WITH THE LOCKING HEAD DURING INSERTION. DO NOT INSERT AT AN ANGLE. AVOID EXCESSIVE FORCE WHEN TIGHTENING IMPLANT. DO NOT USE FORCEPS TO TIGHTEN IMPLANT. DAMAGE RESULTING FROM EXCESSIVE FORCE OR FORCEPS MAY CAUSE IMPLANT FAILURE. PRECAUTIONS: ENSURE THAT THE IMPLANT BODY FOLLOWS THE BONY ANATOMY OF THE STERNUM. SECURE THE LOCKING MECHANISM IN THE INTERCOSTAL SPACE TO MINIMIZE IMPLANT PROFILE. A DIMENSIONAL INSPECTION WAS UNABLE TO BE PERFORMED DUE TO THE POST-MANUFACTURING DAMAGE. A FUNCTIONAL TEST WAS UNABLE TO BE PERFORMED DUE TO THE POST-MANUFACTURING DAMAGE. THE OVERALL COMPLAINT WAS CONFIRMED AS THE OBSERVED CONDITION OF THE STERNAL ZIPFIX CABLE TIE W/NEEDLE PEEK 2 WOULDA HAVE CONTRIBUTED TO THE COMPLAINED DEVICE ISSUE. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE REPORTED COMPLAINT CONDITION. AS PART OF J&J MEDTECH ORTHOPAEDICS QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE HISTORY LOT A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE PRODUCT CODE: 08.501.001.20S LOT NUMBER: 922P606 IT WAS ELECTRONICALLY REVIEWED AND NO NONCONFORMANCES / MANUFACTURING IRREGULARITIES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS. THE PRODUCT WAS RELEASED ON: 20 SEP 2022 MANUFACTURING SITE: I'S A PURCHASED ITEM, SHIPPED BY JABIL BETTLACH AND MANUFACTURED BY SAMAPLAST AG EXPIRY DATE: 01 AUG 2027 IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
Additional Manufacturer Narrative
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H11 ADDITIONAL NARRATIVE: ADDED: B5, D10. CORRECTED: D4 (PRIMARY UDI NUMBER) AND H4. H6 COMPONENT CODES: MOST RELEVANT COMPONENT CODE IS G07002 (APPROPRIATE TERM/CODE NOT AVAILABLE) TO CAPTURE NO FINDINGS AVAILABLE DUE TO NO PRODUCT RETURNED. H3, H4, H6: THE PRODUCT WAS NOT RETURNED TO J&J MEDTECH ORTHOPAEDICS, HOWEVER PHOTOS WERE PROVIDED FOR REVIEW. PHOTOS WERE PROVIDED FOR REVIEW, HOWEVER THE PHOTOS ONLY SHOW THE OPEN PACKAGE AND NO OTHER OBSERVATION PERTAINING TO THE NATURE OF THE REPORTED EVENT COULD BE IDENTIFIED. THEREFORE, THE INVESTIGATION COULD NOT DRAW ANY CONCLUSIONS ABOUT THE REPORTED EVENT DUE TO THE INSUFFICIENT EVIDENCE PROVIDED. IN ADDITION, FUNCTIONALITY OF THE DEVICE CANNOT BE ASSESSED THROUGH PHOTO INVESTIGATION. SINCE THE DEVICE WAS NOT RETURNED, A DIMENSIONAL INSPECTION CANNOT BE PERFORMED. THE OVERALL COMPLAINT WAS NOT CONFIRMED AS THE PHOTOGRAPHS PROVIDED CONTAINED INSUFFICIENT EVIDENCE OF THE REPORTED EVENT. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE COMPLAINT CONDITION. AS PART OF J&J MEDTECH ORTHOPAEDICS QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE HISTORY LOT: A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE. PRODUCT CODE: 08.501.001.20S. LOT NUMBER: 922P606. IT WAS ELECTRONICALLY REVIEWED AND NO NONCONFORMANCES / MANUFACTURING IRREGULARITIES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS. THE PRODUCT WAS RELEASED ON: 20 SEP 2022. MANUFACTURING SITE: I'S A PURCHASED ITEM, SHIPPED BY JABIL BETTLACH AND MANUFACTURED BY SAMAPLAST AG. EXPIRY DATE: 01 AUG 2027. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.
Description of Event or Problem
ADDITIONAL INFORMATION RECEIVED STATES THAT THERE WAS NO SURGICAL DELAY REPORTED AND NO OTHER MEDICAL INTERVENTION WAS REQUIRED. NO BROKEN FRAGMENTS RETAINED IN THE PATIENT.