STERNAL ZIPFIX CABLE TIE W/NEEDLE PEEK 5
Received Nov 14, 2024 · Event occurred Nov 6, 2024
Report 8030965-2024-13940 · MDR key 20685118
Device
Product problems
- Mechanical Problem
- Mechanical Problem
Patient
NA · Unknown
- Insufficient Information
- No Clinical Signs, Symptoms or Conditions
- Insufficient Information
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
ADDITIONAL INFORMATION RECEIVED STATES THAT THE SURGERY WAS COMPLETED SUCCESSFULLY. THERE WAS A SURGICAL DELAY OF 5 MINUTES. PATIENT WAS RECOVERED.
Additional Manufacturer Narrative
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H11 ADDITIONAL NARRATIVE: ADDED: B5, (D4 (EXPIRY DATE), H4 CORRECTED: D4 (PRIMARY UDI NUMBER), H6 (HEALTH EFFECT - IMPACT CODE & HEALTH EFFECT - CLINICAL CODE), G1 H6 COMPONENT CODES: MOST RELEVANT COMPONENT CODE IS G07002 (APPROPRIATE TERM/CODE NOT AVAILABLE) TO CAPTURE NO FINDINGS AVAILABLE DUE TO NO PRODUCT RETURNED. INVESTIGATION SUMMARY: PRODUCT WAS NOT RETURNED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE HISTORY LOT: A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE PRODUCT CODE: 08.501.001.05S LOT NUMBER: 5444P42 IT WAS ELECTRONICALLY REVIEWED AND NO NONCONFORMANCES / MANUFACTURING IRREGULARITIES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS. THE PRODUCT WAS RELEASED ON: 06JUL2023 MANUFACTURING SITE: IS A PURCHASED PRODUCT, RECEIVED FROM SAMAPLAST AG, SHIPPED BY BETTLACH. EXPIRY DATE:31JAN2025 IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
Additional Manufacturer Narrative
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H11 ADDITIONAL NARRATIVE: ADDED: D9 H3, H6: THE DEVICE WAS RECEIVED, THE INVESTIGATION IS IN PROGRESS, NO CONCLUSION COULD BE DRAWN AT THE TIME OF FILING THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.
Description of Event or Problem
IT WAS REPORTED THAT ON (B)(6) 2024, THE EZIPFIX BAND SLIPPED FROM ITS LOCKING HEAD WHICH LEAD TO REMOVAL OF THE 4 BANDS. A NEW PACK WAS USED.
Additional Manufacturer Narrative
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H11 ADDITIONAL NARRATIVE: D4: UDI: AS THE LOT NUMBER FOR THE DEVICE INVOLVED IN THE EVENT WAS NOT PROVIDED, THE FULL UDI IS CURRENTLY NOT AVAILABLE. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
Additional Manufacturer Narrative
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H11 ADDITIONAL NARRATIVE: H3, H4, H6: THE PRODUCT WAS RETURNED TO J&J MEDTECH ORTHOPAEDICS FOR EVALUATION. VISUAL INSPECTION OF THE RETURNED DEVICE FOUND THAT THESTERNAL ZIPFIX CABLE TIE W/NEEDLE PEEK 5 WAS CUT FROM THE LOCKING HEAD AND CUTTING AREA. THE SURFACE FRACTURE INDICATE THAT WAS CUT INTENTIONALLY WITH A CUTTING CLAMP. THIS DAMAGE DEMONSTRATES THAT THE IMPLANT WAS USED. NOT ALL THE CUT FRAGMENTS WERE RETURNED. PER STERNAL ZIPFIX SYSTEM SURGICAL TECHNIQUE GUIDE SE_839363 REV. AB. PRECAUTIONS: DO NOT DAMAGE THE IMPLANT TEETH AND LOCKING HEAD BY MANIPULATING WITH INSTRUMENTS. ENSURE THAT THE LOCKING HEAD OF THE IMPLANT IS FREE OF SOFT TISSUE AND/OR SURGICAL MATERIAL THAT COULD PREVENT LOCKING OF THE IMPLANT. AVOID CLAMPING OF IMPLANT IN THE AREA OF THE TEETH OR EXCESSIVE BENDING/TWISTING OF THE IMPLANT, AS THIS MAY LEAD TO IMPLANT FAILURE. DO NOT CUT THE IMPLANT DIRECTLY AT THE NOTCH. REMOVING THE NEEDLE BY BENDING OR TWISTING WILL CAUSE A DEFORMED END THAT MAY DAMAGE THE LOCKING HEAD DURING INSERTION. ALWAYS ENSURE THAT THE IMPLANT END IS CUT AND NOT DEFORMED. IF THE IMPLANT IS NOT CUT, IMPLANT FAILURE MAY OCCUR. HANDLE IMPLANTS CAREFULLY, ESPECIALLY NEEDLES, TO AVOID DAMAGING CRITICAL STRUCTURES, SOFT TISSUE AND/OR HAND GLOVES. AVOID EXCESSIVE FORCE WHEN TIGHTENING IMPLANT. DO NOT USE FORCEPS TO TIGHTEN IMPLANT. DAMAGE RESULTING FROM EXCESSIVE FORCE OR FORCEPS MAY CAUSE IMPLANT FAILURE. A FUNCTIONAL TEST WAS UNABLE TO BE PERFORMED DUE TO THE POST-MANUFACTURING DAMAGE. HOWEVER, BASED ON THE INVESTIGATION EVIDENCE AND THE EVENT DESCRIPTION, THE COMPLAINT CAN BE CONFIRMED. NEVERTHELESS, IS NOT POSSIBLE TO DETERMINE A POTENTIAL CAUSE AT THIS MOMENT. THE MODE OF FAILURE OF THE DEVICE IS MULTI-FACTORIAL AND CONSIDERATION MUST BE GIVEN TO ALL OTHER POTENTIAL INFLUENCES SUCH AS SURGICAL PROCESS, PATIENT VARIABLES (E.G. ACTIVITY LEVEL AND USE), ANATOMICAL CONSIDERATIONS AND PATIENT CHANGES OVER TIME. ANY CONCLUSIONS FROM THE INVESTIGATIONAL INPUT PROVIDED HAVE TO BE PLACED INTO CONTEXT WITH ALL OTHER RELEVANT FACTORS. A DIMENSIONAL INSPECTION WAS NOT PERFORMED SINCE IT WAS NOT APPLICABLE TO THE COMPLAINT CONDITION. THE OVERALL COMPLAINT WAS CONFIRMED AS THE OBSERVED CONDITION OF THE STERNAL ZIPFIX CABLE TIE W/NEEDLE PEEK 5 WOULD HAVE CONTRIBUTED TO THE COMPLAINED DEVICE ISSUE. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE REPORTED COMPLAINT CONDITION. AS PART OF J&J MEDTECH ORTHOPAEDICS QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE HISTORY REVIEW A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE PRODUCT CODE: 08.501.001.05S LOT NUMBER: 5444P42 IT WAS ELECTRONICALLY REVIEWED AND NO NONCONFORMANCES / MANUFACTURING IRREGULARITIES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS. THE PRODUCT WAS RELEASED ON: 06JUL2023 MANUFACTURING SITE: IS A PURCHASED PRODUCT , RECEIVED FROM SAMAPLAST AG, SHIPPED BY BETTLACH. EXPIRY DATE: 31JAN2025.