CD HORIZON® FENESTRATED SCREW SET
Received Nov 13, 2024 · Event occurred Oct 15, 2024
Report 1030489-2024-01489 · MDR key 20672164
Device
Generic name
Bone Cement, Posterior Screw Augmentation
Manufacturer
Mdt Sofamor Danek Puerto Rico MfgModel number
55751027545Catalog number
55751027545
Lot number
H5946764
Product problems
- Use of Device Problem
- Unintended Movement
- Use of Device Problem
- Unintended Movement
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
H3: PRODUCT ANALYSIS OF PART # 55751027545; LOT # H5946764 VISUAL INSPECTION DID NOT REVEAL ANY DAMAGE TO THE SCREW. A DIFFERENT SIZE SCREW WAS USED FOR THE PROCEDURE. NO FAULT FOUND. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
Description of Event or Problem
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT UNDERGOING POSTERIOR F USION SPINAL THERAPY. IT WAS REPORTED THAT THE SCREW WAS LOOSE. THERE WAS NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT. ADDITIONAL INFORMATION RECEIVED FROM MANUFACTURER REPRESENTATIVE THAT THE DEVICE HAS NOT RECEIVED AS DEFECT PRODUCT AND NOT A MANUFACTURING ISSUE. THE HOLE WAS CREATED ACCORDING TO THE SUITABLE DIAMETER OF SCREW. THE PILOT HOLE'S DIAMETER WAS SMALLER THAN THAT OF THE SCREW.
Additional Manufacturer Narrative
G2: COUNTRY OF ORIGIN IS JAPAN. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.