NA ELECTRODE
Received Nov 12, 2024 · Event occurred Aug 31, 2024
Report 1823260-2024-03274 · MDR key 20666805
Device
Generic name
Electrode, Ion Specific, Sodium
Manufacturer
Roche DiagnosticsCatalog number
10825468001
Lot number
ASKU
Product problems
- High Test Results
- Low Readings
- High Test Results
- Low Readings
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THE ANALYZER'S SERIAL NUMBER IS (B)(6). THE ELECTRODES WERE REPLACED, WASH RACKS WERE RERUN, AND CALIBRATION AND QC WERE PERFORMED. THE INVESTIGATION IS ONGOING.
Description of Event or Problem
THERE WAS AN ALLEGATION OF QUESTIONABLE NA ELECTRODE RESULTS FOR 37 PATIENT SAMPLES ON A COBAS PRO ISE ANALYTICAL UNIT. PLEASE REFER TO THE ATTACHMENT "(B)(6)" FOR EXAMPLES OF QUESTIONABLE PATIENT RESULTS. THE QUESTIONABLE RESULTS WERE REPORTED OUTSIDE THE LABORATORY AND CORRECTED REPORTS WERE SENT. THE REPEATED RESULTS WERE OBTAINED ON ANOTHER MODULE AND WERE BELIEVED TO BE CORRECT. THE SAMPLES WERE REPEATED DUE TO FAILED QUALITY CONTROLS.
Additional Manufacturer Narrative
CALIBRATION AND QC WERE ACCEPTABLE. A GENERAL REAGENT ISSUE WAS EXCLUDED. THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM.