inforMED
MalfunctionMRZ

CADD-SOLIS COMM MODULE

Received Nov 11, 2024 · Event occurred Oct 1, 2024

Report 3012307300-2024-13115 · MDR key 20658571

Device

Generic name

Accessories, Pump, Infusion

Model number

21-2131-25

Catalog number

21-2131-25

Lot number

10343942

Product problems

  • Failure to Infuse
  • Wireless Communication Problem
  • Failure to Infuse
  • Wireless Communication Problem

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

B3: UNKNOWN; NO INFORMATION HAS BEEN PROVIDED TO DATE. H3: THE DEVICE WAS RECEIVED FOR EVALUATION. SERVICE HISTORY REVIEW IDENTIFIED THERE WAS NO INDICATION THAT THE COMPLAINT WAS RELATED TO A SERVICE OF THE DEVICE WITHIN THE REVIEW PERIOD. VISUAL AND FUNCTIONAL TESTS WERE PERFORMED WHICH FOUND A DEPLETED BATTERY STATUS AND THE COMM MODULE FAILED TO ENABLE THE WIRELESS SETTINGS OF THE PUMP AND CONNECT TO THE WIRELESS NETWORK. THE CAUSE WAS SUSPECTED TO BE THE WIRELESS RADIO BOARD. THE COMM MODULE IS BEING SENT TO SUPPLIER FOR FURTHER ANALYSIS.

Description of Event or Problem

IT WAS REPORTED THAT THE WIRELESS COMMUNICATION MODULE WAS NOT WORKING (OUT OF BOX). THE RESULTS OF THE INVESTIGATION FOUND THAT THERE WAS A DEPLETED BATTERY STATUS. THERE WAS NO PATIENT INVOLVEMENT, AND NO ADVERSE EVENT REPORTED.