inforMED
MalfunctionQOF

BD RESPIRATORY VIRAL PANEL FOR BD MAX¿ SYSTEM

Received Nov 6, 2024 · Event occurred Oct 11, 2024

Report 3007420875-2024-00117 · MDR key 20620290

Device

Generic name

Multiple-genus Respiratory Virus Nucleic Acid Ivd, Kit, Nucleic Acid Technique

Catalog number

445215

Lot number

4194293

Product problems

  • False Positive Result
  • False Positive Result

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

D2: ADDITIONAL MEDICAL DEVICE TYPES: QQX. E1: INITIAL REPORTER PHONE NUMBER: (B)(6). A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem

REPORT 4 OF 4: IT WAS REPORTED THAT DURING USE OF THE BD RESPIRATORY VIRAL PANEL FOR BD MAX¿ SYSTEM, A FALSE POSITIVE FLU A PATIENT RESULT WAS OBTAINED. THERE WAS NO REPORT OF PATIENT IMPACT.

Description of Event or Problem

REPORT 4 OF 4: IT WAS REPORTED THAT DURING USE OF THE BD RESPIRATORY VIRAL PANEL FOR BD MAX¿ SYSTEM, A FALSE POSITIVE FLU A PATIENT RESULT WAS OBTAINED.

Additional Manufacturer Narrative

INVESTIGATION SUMMARY: THERE IS AN INDICATION OF A BD PCR CARTRIDGES (REF. (B)(4)) ISSUE FOR THE LOT USED BY THE CUSTOMER BASED ON THE ANALYSIS OF THE COMPLAINTS RECEIVED FOR UNUSUAL CURVES PRESENTING AN UPWARD DRIFT IN THE CY5 CHANNEL WHEN USING THE BD PCR CARTRIDGES WITH THE BD MAX RVP ASSAY. THE ROOT CAUSE FOR THE CY5 SIGNAL DRIFT ISSUE ASSOCIATED WITH THE BD PCR CARTRIDGES LOT WAS NOT IDENTIFIED BUT A CORRECTIVE AND PREVENTIVE ACTION HAS BEEN INITIATED TO IDENTIFY ROOT CAUSE. ALTHOUGH THE ISSUE WAS NOT REPRODUCED WHEN TESTING RETAIN CARTRIDGE MATERIAL, BD CONFIRMS THE COMPLAINT BASED ON THE INVESTIGATION THAT WAS PERFORMED.BD QUALITY WILL CONTINUE TO MONITOR FOR TRENDS.