Received Nov 6, 2024 · Event occurred Oct 21, 2024

Report 9616086-2024-00101 · MDR key 20616744

Device

Generic name

X-rom Post-op Knee Brace

Manufacturer

Djo Llc

Model number

11-2181-9

Lot number

UNKNOWN

Product problems

  • Mechanical Problem
  • Unintended System Motion
  • No Apparent Adverse Event
  • Mechanical Problem
  • Unintended System Motion
  • No Apparent Adverse Event

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT ONE OF THE LOCKING MECHANISMS NO LONGER HOLDS AND THE HINGE WILL RANDOMLY UNLOCK. PATIENT HAD THE BRACE FOR 2 WEEKS BEFORE IT STOPPED WORKING. (PROBLEM WITH RED BUTTON)

Additional Manufacturer Narrative

ONE OF THE LOCKING MECHANISMS NO LONGER HOLDS AND THE HINGE WILL RANDOMLY UNLOCK. PATIENT HAD THE BRACE FOR 2 WEEKS BEFORE IT STOPPED WORKING. (PROBLEM WITH RED BUTTON). THE BRACE HAS NOT BEEN RETURNED FOR EVALUATION. THE ROOT CAUSE OF THE COMPLAINT IS DAMAGED COMPONENT BASED ON OTHER DEVICES THAT HAVE BEEN EVALUATED. THE DEVICES LOCK BUTTON FAILED DURING MANUFACTURER TESTING CONFIRMED TO BE THE SAME MALFUNCTION FOUND IN REPORT NUMBER 9616086-2023-00007. ADDITIONAL REPORTING ON THIS EVENT WILL BE PROVIDED AS A SUPPLEMENTAL REPORT TO THIS DOCUMENT IF IT BECOMES AVAILABLE.